- 2025-05-21FDA-DeviceClass IIPiezoWave 2 Control Unit, Product Number 100506US, compact myofascial acoustic compression therapy device that enables treatment with piezoelectric shockwaves (Therapeutic massager).
The high voltage capacitor may fail early in the product life cycle after delivery of surge voltages. When this failure occurs, the capacitor may produce smoke and a burning smell. In extremely rare instances a capacitor may cause a spark or excessive heat when an internal capacitor failure occurs.
- 2023-05-03FDA-DeviceClass IIFLEX. GRASP. FORCEPS 6.6FR WL 550MM, Product Number 8736.685
Potentially unable to open forceps jaws when the endoscope shaft is bent by about 90 degrees. If, in addition, the proximal forceps shaft is bent, the forceps can no longer be opened or restrictions begin even with a slight bending of the endoscope.
- 2023-05-03FDA-DeviceClass IIFLEX. GRASP. FORCEPS 5FR WL 550MM, Product Number 8735.685
Potentially unable to open forceps jaws when the endoscope shaft is bent by about 90 degrees. If, in addition, the proximal forceps shaft is bent, the forceps can no longer be opened or restrictions begin even with a slight bending of the endoscope.
- 2023-05-03FDA-DeviceClass IIFLEX. BIOPSY FORCEPS 5FR WL 550MM, Product Number 829.051
Potentially unable to open forceps jaws when the endoscope shaft is bent by about 90 degrees. If, in addition, the proximal forceps shaft is bent, the forceps can no longer be opened or restrictions begin even with a slight bending of the endoscope.
- 2023-05-03FDA-DeviceClass IIFLEX. GRASP. FORCEPS 5FR WL 550MM, Product Number 828.051
Potentially unable to open forceps jaws when the endoscope shaft is bent by about 90 degrees. If, in addition, the proximal forceps shaft is bent, the forceps can no longer be opened or restrictions begin even with a slight bending of the endoscope.