FDA DeviceClass II

Medline kits containing Tego Connectors Medline APHERESIS SUBQPORT ACCESS KIT SKU DYNDA2705 Medline CENTRAL LINE INSERTION SKU DYNJ63347A

Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.

From the FDA notice · Read the official notice ↗ · All Medline recalls

What to do

The agency's notice has the remedy and who to contact.

Is yours affected?

Compare your label or packaging with the recalled product.

Product
Medline kits containing Tego Connectors Medline APHERESIS SUBQPORT ACCESS KIT SKU DYNDA2705 Medline CENTRAL LINE INSERTION SKU DYNJ63347A
Sold in
Nationwide
Made for
All ages (under 100 years)

This recall covers 6 products — see all of them.

Details

Units affected
1212
Illnesses reported
None reported
Manufactured in
3 Lakes Dr, N/A, Northfield, IL, United States
Classification
Class II
Recall number
Z-1466-2026

In the FDA's words

Description
Medline kits containing Tego Connectors Medline APHERESIS SUBQPORT ACCESS KIT SKU DYNDA2705 Medline CENTRAL LINE INSERTION SKU DYNJ63347A Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body. SKU DYNDA2705 UDI-DI (ea) 10193489468632 UDI-DI (case) 40193489468633 Kit lots 24DBL787 24HBR630 25ABH699 25DBV209 25GBQ405 SKU DYNJ63347A UDI-DI (ea) 10193489278064 UDI-DI (case) 40193489278065 Kit lots 24AMC874 24AME139 24AMH820 24BMF097 24CMF691 24GMB769 24GMK036 24IMC124 24LMA110 25AMJ634 25GMD386 25IMB510 25JMI708 Distribution: US Nationwide distribution. Affected quantity: 1212
Hazard
Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.

6 products in this recall

Medline kits with Tego Connectors recalled for seal failure risk

Medline is recalling six kits that include Tego Connectors, including dialysis, central line, and dressing change kits from Medline and Centurion. Affected lots may have silicone seal doming or tearing, which can block fluid flow, delay or interrupt therapy, cause leakage, or allow air or biological contaminants into the body. No injuries have been reported.