Agiliti Health Adapt Pump recalled for temperature control malfunction
Agiliti Health's Adapt Pump (used in Adapt Air and Adapt Convertible mattresses) has a software issue that prevents the temperature control (Microclimate Management) function from restarting after the Autofirm button is used for 10 minutes during initial setup. This only happens if you let the timer run out without manually stopping it, and it only affects temperature management—not the mattress's air support.
From the FDA notice · Read the official notice ↗
What to do
- Check if your mattress pump was made before December 4, 2025 (see the UDI label: 0084569904914)
- If affected, avoid using the Autofirm function alone during setup, or manually press Autofirm again before the 10-minute timer ends
- Contact Agiliti Health for a software update or replacement pump
Is yours affected?
Compare your label or packaging with the recalled product.
- Product
- Adapt Pump, Model/Catalog Number: 61600200-Adapt Pump, component of Adapt support surfaces (Adapt Air, Adapt Air EX, Adapt Air Pro, Adapt Air Pro EX, Adapt Convertible, Adapt Convertible EX), sofware version 1.3.0; Mattress, Air Flotation, Alternating Pressure
- Sold in
- Nationwide
- Made for
- Adults (0 months and older)
Details
- Units affected
- 4286
- Illnesses reported
- None reported
- Manufactured in
- 204 W 2nd St, N/A, Ellis, KS, United States
- Classification
- Class II
- Recall number
- Z-1472-2026
In the FDA's words
- Description
- Adapt Pump, Model/Catalog Number: 61600200-Adapt Pump, component of Adapt support surfaces (Adapt Air, Adapt Air EX, Adapt Air Pro, Adapt Air Pro EX, Adapt Convertible, Adapt Convertible EX), sofware version 1.3.0; Mattress, Air Flotation, Alternating Pressure Agiliti identified an issue where the Adapt Pump (61600200-Adapt Pump) supplied with all Adapt Line support surfaces does not re-engage the Microclimate Management function after using the Autofirm function for 10 minutes. This issue only occurs when the 10-minute timer is allowed to elapse after the Autofirm button is pressed after initial setup, and it only affects the Microclimate Management function. UDI-DI 0084569904914; All serial numbers with a manufacturing date prior to 12/04/2025 Distribution: US Nationwide distribution. Affected quantity: 4286
- Hazard
- Agiliti identified an issue where the Adapt Pump (61600200-Adapt Pump) supplied with all Adapt Line support surfaces does not re-engage the Microclimate Management function after using the Autofirm function for 10 minutes. This issue only occurs when the 10-minute timer is allowed to elapse after the Autofirm button is pressed after initial setup, and it only affects the Microclimate Management function.