Philips Xper Flex Cardio (FC2010, Procedure Room and FC2020, Bedside Monitor) devices - intended for use by professional healthcare providers for complete physiologic/hemodynamic monitoring Service numbers associated with FC2010: Xper Flex Cardio 2010 Rev A 453564241901 Xper Flex Cardio 2010 Rev B 453564483321 Flex Cardio FC2010 Rev C 453564621791 Xper Flex Cardio 2010 REV C RUEXCHANGE 453564675021 Xper Flex Cardio 2010 Rev D 453564669081 Xper Flex Cardio 2010 Rev D Russia 453564674581
Performance issues with the Xper Flex Cardio Physio Monitoring System include: potential delay of up to 10 seconds in displaying ECG, invasive blood pressure and other parameters; patient weight is rounded to the nearest whole kilogram;Xper IM software used with the Xper Flex Cardio Physio Monitoring System may periodically crash; No SpO2 numeric or plethysmography is displayed when SpO2 is connected o the Flex Cardio device; display of certain data from the FC2010 becomes frozen, i.e., waveforms cease sweeping and updating and the ECG, IBP, and respiration numeric values cease to update;e ECG, IBP, and respiration waveforms become flat lines and no audible alarms are produced for HR and IBP, upon start up, an unexpected non physiological ECG waveform, erratic heart rate numeric value, and non-physiological display of any other active waveforms may appear on the Boom Monitor
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1517-2021Don't want to check this manually?