FDA-Device2026-06-17Class IIPACKAGING DEFECT

Philips DigitalDiagnost C90 Flex/Value/Chest/ER recalled for missing seismic stabilization kit

What this means

One Philips DigitalDiagnost C90 Flex/Value/Chest/ER X-ray system was shipped without the required Seismic kit for its Wall Stand VS2. The kit is designed to keep the stand stable and prevent damage during earthquakes or seismic events.

  • If you operate a Philips DigitalDiagnost C90 with model number 712035, contact Philips to obtain the missing Seismic kit for the Wall Stand VS2.
  • Do not use the Wall Stand VS2 in an earthquake-prone area until the seismic kit is installed.
The agency's original wording
Hazard

System was delivered without the required Seismic kit for Wall Stand VS2, which is designed to prevent damage and ensure stability of the Wall Stand VS2 during seismic events.

Intended for
All ages
Illnesses reported
None reported
Sold states
US distribution to California.
Affected count
1 unit
Manufactured in
222 Jacobs St, N/A, Cambridge, MA, United States
Products
DigitalDiagnost C90 Flex/Value/Chest/ER. Model Number: 712035.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2354-2026