FDA-Device2022-08-31Class I

HAMILTON-C6 Intensive Care Ventilator, REF: 160021

Hamilton Medical AG
Hazard

Ventilator status indicator board can become loose, which could lead to water ingress (disinfectants) that may lead to technical fault alarms. Multiple technical faults in a short time may force ventilators into Safety Ventilation (blower runs constantly) or Ambient Sate (inspiratory channel/expiratory valves opened; patient breaths room air unassisted) with Panel connection lost message displayed

Sold states
US nationwide distribution including Puerto Rico.
Affected count
358
Manufactured in
Via Crusch 8, Bonaduz, Switzerland
Products
HAMILTON-C6 Intensive Care Ventilator, REF: 160021

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1525-2022

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief