Serious riskFDA DeviceClass I

Hamilton Medical circuit sets recalled for possible valve sticking

Certain coaxial breathing circuit sets used with Hamilton Medical ventilators (models C1, T1, and MR1) may have expiratory valve membranes that stick to the valve body. This defect is not always detectable before use and can trigger an "exhalation obstructed" alarm during ventilation, potentially leading to inadequate ventilation and oxygen desaturation in patients.

From the FDA notice · Read the official notice ↗

What to do

  1. Perform a lung action test on the breathing circuit prior to each use
  2. If an "exhalation obstructed" alarm occurs during use, remove and inspect the expiratory valve set or replace it immediately
  3. Contact Hamilton Medical if you have questions about affected lot numbers (200379–205050)

Is yours affected?

Compare your label or packaging with the recalled product.

Product
Breathing circuit set, coaxial: REF: 260127, 260128, 260167, 260168 used with HAMILTON-C1, HAMILTON-T1 and HAMILTON-MR1 ventilators.
Sold in
Nationwide
Made for
All ages (under 100 years)

Details

Units affected
27,270 pcs
Illnesses reported
None reported
Manufactured in
Parc Industrial Vial 10, N/A, Domat/Ems, N/A, Switzerland
Classification
Class I
Recall number
Z-2463-2026

In the FDA's words

Description
Breathing circuit set, coaxial: REF: 260127, 260128, 260167, 260168 used with HAMILTON-C1, HAMILTON-T1 and HAMILTON-MR1 ventilators. Ventilator coaxial breathing circuit may have, expiratory valve sets with membranes sticking to valve body, not detectable during pre-operative test, which will result in "exhalation obstructed " alarm, which may result in inadequate ventilation, oxygen desaturation. Advising: Prior to use, test lung action. If alarm occurs, expiratory valve set removal to overcome membrane adhesion or replacement REF/UDI-DI: 260127/07630002802956, 17630002802953; 260128/07630002802963, 17630002802960; 260167/07630002802970, 17630002802977; 260168/07630002802987, 17630002802984. Lot numbers between 200379 and 205050 (including both.) Distribution: US Nationwide distribution including in the states of KS, WA, VT, TX, WI, AR, OH, PA, CA, CO, ID, AZ, ME, MO, MI, IL, OR, VA, FL, NJ, SD, NY, NM, TN, CT, NE, OK, AL, GA, KY, NH, WY, ND, IN, UT, WV, DE, IA, PR, MN, LA, AK, HI. Affected quantity: 27,270 pcs
Hazard
Ventilator coaxial breathing circuit may have, expiratory valve sets with membranes sticking to valve body, not detectable during pre-operative test, which will result in "exhalation obstructed " alarm, which may result in inadequate ventilation, oxygen desaturation. Advising: Prior to use, test lung action. If alarm occurs, expiratory valve set removal to overcome membrane adhesion or replacement