Physio-Control infant/child defibrillator electrodes recalled for delamination risk
Physio-Control infant and child defibrillator electrodes may separate or peel apart during use, which could prevent proper delivery of therapy. The affected electrodes are pre-gelled, self-adhesive pads designed for use with automated external defibrillators (AEDs) on children under 8 years old or weighing less than 55 pounds. No injuries have been reported.
From the FDA notice · Read the official notice ↗
What to do
- Check your AED equipment for the affected catalog numbers: 11101-000016 or 11101-000017
- If you have these electrodes, stop using them immediately and remove them from your AED
- Contact Physio-Control or your AED supplier for replacement electrodes
- Verify your replacement electrodes are from an unaffected lot number
Is yours affected?
Compare your label or packaging with the recalled product.
- Product
- PHYSIO CONTROL INFANT/CHILD Reduced Energy Defibrillator Electrodes Catalog Numbers: 11101-000016 UDI-DI code: 00721902629013 Catalog Number: 11101-000017 UDI-DI code: 00721902682483 Infant/Child Reduced Energy Defibrillation Electrodes are pre gelled, self-adhesive, therapy electrodes that allow hands free defibrillation. These electrodes reduce the energy delivered to the patient by a factor of 4:1. The Infant/child reduced energy defibrillation electrodes are therapy electrodes designed to allow personnel who are trained on AED device operation and in basic life support, or other physician authorized emergency medical response systems, to safely and effectively defibrillate infants and children who are less than 8 years old or weigh less than 25kg (55lbs).
- Sold in
- Nationwide
- Made for
- Children (under 8 years)
Details
- Units affected
- 16,499 electrodes (US/OUS: 8327/8172)
- Illnesses reported
- None reported
- Manufactured in
- 11811 Willows Rd Ne, N/A, Redmond, WA, United States
- Classification
- Class II
- Recall number
- Z-1544-2026
In the FDA's words
- Description
- PHYSIO CONTROL INFANT/CHILD Reduced Energy Defibrillator Electrodes Catalog Numbers: 11101-000016 UDI-DI code: 00721902629013 Catalog Number: 11101-000017 UDI-DI code: 00721902682483 Infant/Child Reduced Energy Defibrillation Electrodes are pre gelled, self-adhesive, therapy electrodes that allow hands free defibrillation. These electrodes reduce the energy delivered to the patient by a factor of 4:1. The Infant/child reduced energy defibrillation electrodes are therapy electrodes designed to allow personnel who are trained on AED device operation and in basic life support, or other physician authorized emergency medical response systems, to safely and effectively defibrillate infants and children who are less than 8 years old or weigh less than 25kg (55lbs). Due to pediatric defibrillator electrode delamination Catalog Numbers: 11101-000016 UDI-DI code: 00721902629013 Catalog Number: 11101-000017 UDI-DI code: 00721902682483 Lot Numbers: 315303, 315639, 315858, 316007, 320529, 318931 Distribution: Worldwide - U.S. Nationwide distribution including in the states of AK, AL, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV. The countries of Australia, Canada, Hong Kong, Japan, Mexico, Malaysia, Netherlands, New Zealand, and Taiwan. Affected quantity: 16,499 electrodes (US/OUS: 8327/8172)
- Hazard
- Due to pediatric defibrillator electrode delamination