EndoVive Safety PEG Kit; Outer Box Material Number (UPN): M00509001; Inner Box Material Number (UPN): M00509000;
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1551-2026Original summary
EndoVive Safety PEG Kit; Outer Box Material Number (UPN): M00509001; Inner Box Material Number (UPN): M00509000; Recall of kits which utilized ChloraPrep Triple Swabsticks from BD. Open or compromised seals on the sterile swabstick packaging, which could result in loss of sterility or drying of the antiseptic solution. This component is placed outside of the sterile barrier of the tray manufactured by Boston Scientific, which contains the majority of the kit components. As such, inclusion of the affected ChloraPrep Triple Swabsticks does not degrade the integrity of the other components within the PEG kit. Outer Box Material Number (UPN): M00509001; Inner Box Material Number (UPN): M00509000; Outer Box UDI: 08714729880394; Inner Box UDI: 08714729880387; Outer Box Lot#s: 37435463, 37522541, 37528545, 37546847, 37546850, 37546849, 37550500, 37592036, 37732292; Inner Box Lot#s: 37435462, 37514809, 37525231, 37536563, 37536564, 37536562, 37546840, 37583642, 37731897; Distribution: US Nationwide distribution. Affected quantity: 806 units
Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief