FDA-Device2024-04-24Class II
Siemens Artis Models: Interventional Fluoroscopic X-Ray Systems-Models: Artis zee floor- 10094135 Artis zee ceiling- 10094137 Artis zee multi-purpose- 10094139 Artis zee biplane- 10094141 Artis zeego- 10280959 Artis Q floor- 10848280 Artis Q ceiling- 10848281 Artis Q biplane- 10848282 Artis Q zeego- 10848283 Artis Q.zen floor- 10848353 Artis Q.zen ceiling- 10848354 Artis Q.zen biplane- 10848355 Artis zee floor MN- 10094142 Artis zee biplane MN- 10094143
Siemens Medical Solutions USA, Inc
Hazard
Coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message TUBE HOT, have a break . the system cannot be operated normally, may result in a situation where it is necessary to cancel clinical treatment or to continue treatment on an alternative system
Sold states
US Nationwide distribution.
Affected count
583 units
Manufactured in
40 Liberty Blvd, Malvern, PA, United States
Products
Siemens Artis Models: Interventional Fluoroscopic X-Ray Systems-Models: Artis zee floor- 10094135 Artis zee ceiling- 10094137 Artis zee multi-purpose- 10094139 Artis zee biplane- 10094141 Artis zeego- 10280959 Artis Q floor- 10848280 Artis Q ceiling- 10848281 Artis Q biplane- 10848282 Artis Q zeego- 10848283 Artis Q.zen floor- 10848353 Artis Q.zen ceiling- 10848354 Artis Q.zen biplane- 10848355 Artis zee floor MN- 10094142 Artis zee biplane MN- 10094143
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1598-2024Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief