FDA-Device2024-05-01Class II
ACUSON Maple Diagnostic Ultrasound System, REF: 11711750, ACUSON Maple Select, REF: 11711779, running software version VA10A
Siemens Medical Solutions USA, Inc.
Hazard
On ultrasound systems, when Cardiac DICOM SR feature is configured to display either minimum or maximum measured value, and multiple cardiac region measurements are made, and results are exported into the SR feature, then The SR viewer will display the LAST measured value, not min or max, which could contribute to patient condition misdiagnosis or negatively influence patient management decisions.
Sold states
Worldwide distribution - US Nationwide including in the states of AL, AR, CA, CO , CT, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA , MD, MI, MN, MO, MS, NC, ND, NE, NJ, NM, NY, OH, OK, OR, PA, PR, RI, SD, TN, TX, UT, VA, WA, WI, WY and the countries of BG, BW, BA, TR, AR, CL, CN, FR, SK, CD, BH, PH, AM, CR, AF, BE, AE, AZ, AO, KR, RO, HU, HR, DK, AU, DZ, CY, PL, BO, AL, NL, CI, BR, CZ, DE, AT, ES, CO, BD, CH, IT, CA, CM, DO, EC, EG, ER, ET, FI, FO, GB, GE, GR, GT, GY, HK, HN, ID, IE, IL, IN, IQ, IR, JO, JP, KE, KW, KZ, LB, LT, LU, LV, LY, MA, MD, ME, MK, MM, MN, MU, MW, MX, MY, NG, NO, NP, NZ, OM, PA, PE, PG, PK, PT, PY, QA, RS, RU, SA, SE, SG, SI, SM, SN, SV, SZ, TH, TN, TT, TW, TZ, UA, UG, UY, UZ, VN, XK, YE, ZA, ZM, ZW.
Affected count
67
Manufactured in
22010 Se 51st St, N/A, Issaquah, WA, United States
Products
ACUSON Maple Diagnostic Ultrasound System, REF: 11711750, ACUSON Maple Select, REF: 11711779, running software version VA10A
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1601-2024Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief