Custom Abutment AS Ti, for Straumann RC, incl. TAN screw; Catalog Number/Art.: 010.5438;
A material mix-up had occurred during the manufacturing of Straumann¿ Custom Abutments. More specifically, a raw material bar of TAN-material (SAP #701924 / Lot 7130439) was incorrectly identified as a Titanium Bar Ti Gr4 D (SAP #701922 / Lot 7130430). As a result, RC Titanium abutments were manufactured using the TAN Bar material on Mill-from-Bar Machine 396-06.
From the FDA notice · Read the official notice ↗ · All Straumann recalls
What to do
The agency's notice has the remedy and who to contact.
Is yours affected?
Compare your label or packaging with the recalled product.
- Product
- Custom Abutment AS Ti, for Straumann RC, incl. TAN screw; Catalog Number/Art.: 010.5438;
- Sold in
- Nationwide
- Made for
- Adults (0 months and older)
This recall covers 2 products — see all of them.
Details
- Units affected
- 7 units
- Illnesses reported
- None reported
- Manufactured in
- 60 Minuteman Rd, Andover, MA, United States
- Classification
- Class II
- Recall number
- Z-1633-2026
In the FDA's words
- Description
- Custom Abutment AS Ti, for Straumann RC, incl. TAN screw; Catalog Number/Art.: 010.5438; A material mix-up had occurred during the manufacturing of Straumann¿ Custom Abutments. More specifically, a raw material bar of TAN-material (SAP #701924 / Lot 7130439) was incorrectly identified as a Titanium Bar Ti Gr4 D (SAP #701922 / Lot 7130430). As a result, RC Titanium abutments were manufactured using the TAN Bar material on Mill-from-Bar Machine 396-06. Catalog Number/Art.: 010.5438; UDI-DI: 07630031773937; Order number (lot): 71315US_00729_01_b00, 91076US_03269_00_b00, 72509US_01048_00_b00, 72839US_00065_00_b00, 39133US_11508_02_b00, 4202CA_00558_00_b00, 72630CA_00109_00_b00; Distribution: Worldwide - US Nationwide distribution in the states of CA, CO, FL, IA, IL, MN, MO, MT, OR, WA; and the country of Canada. Affected quantity: 7 units
- Hazard
- A material mix-up had occurred during the manufacturing of Straumann¿ Custom Abutments. More specifically, a raw material bar of TAN-material (SAP #701924 / Lot 7130439) was incorrectly identified as a Titanium Bar Ti Gr4 D (SAP #701922 / Lot 7130430). As a result, RC Titanium abutments were manufactured using the TAN Bar material on Mill-from-Bar Machine 396-06.
2 products in this recall
Straumann custom abutments recalled for wrong raw material
Straumann is recalling two custom abutment products, 20 units in total, for dental implant use. A raw material bar of TAN material was mislabeled as titanium Grade 4, so some RC Titanium abutments were made from the wrong material. The affected units were distributed in several U.S. states and in Canada. No injuries have been reported.
The agency lists each product separately. Check every one you might have.
- Custom Abutment Ti, for Straumann RC, incl. TAN screw; Catalog Number/Art.: 027.4620;Z-1632-2026 · 13 units
- Custom Abutment AS Ti, for Straumann RC, incl. TAN screw; Catalog Number/Art.: 010.5438; (this page)Z-1633-2026 · 7 units