FDA-Device2025-04-30Class II

Skytron Freedom 4FXS-60 Heavy Duty Monitor Mount

IHB OPERATIONS B.V.
Hazard

The mounting hardware that is supplied with the 4FXS-60 monitor bracket does not provide adequate thread engagement to support the weight of the frame and monitor. This could lead to a failure of the mounting hardware and subsequent separation of the frame from the mounting column.

Sold states
US Nationwide distribution including in the states of AL, AZ, CA, FL, IA, ID, IL, LA, MA, MD, MI, MN, NC, NH, NJ, NM, NY, PA, TN, TX, UT, VA, WI, WV.
Affected count
112 units
Manufactured in
Floresstraat 52, N/A, Zwolle, N/A, Netherlands
Products
Skytron Freedom 4FXS-60 Heavy Duty Monitor Mount

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1663-2025