- 2025-04-30FDA-DeviceClass IISkytron Freedom 4FXS-60 Heavy Duty Monitor Mount
The mounting hardware that is supplied with the 4FXS-60 monitor bracket does not provide adequate thread engagement to support the weight of the frame and monitor. This could lead to a failure of the mounting hardware and subsequent separation of the frame from the mounting column.
- 2023-12-13FDA-DeviceClass IIFreedom Boom models series: F110, F120, F130, F320, F330, F350, F410, and F420 with built in 42A Lightweight Spring Arm part numbers 42A-43L, 42A-43M, 42A-43H, and 42A-43S. Ceiling mounted utilities and equipment dispensing system.
Weakened connection point on the adjustable suspension arm, potentially causing damage or a break.