FDA-Device2026-04-15Class IPROCESSING DEFECT

Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: CVOR ANGIOGRAPHY PACK-LF, Medline SKU DYNJ42367B

OTHERNationwide distribution
Hazard

Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.

Sold states
Worldwide distribution: US (nationwide) including Puerto Rico (PR) and OUS (foreign) to countries of: Canada (CA), Netherlands (NL), Australia (AU), Korea, Republic of (KR), Sri Lanka (LK), Pakistan (PK), Japan (JP), United Arab Emirates (AE), Singapore (SG) and Slovakia (SK).
Affected count
3120 kits
Manufactured in
3 Lakes Dr, Northfield, IL, United States
Products
Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: CVOR ANGIOGRAPHY PACK-LF, Medline SKU DYNJ42367B

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1714-2026

More Medline Industries, LP recalls

Don't want to check this manually?