FDA-Device2026-04-15Class IPROCESSING DEFECT
Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: CVOR ANGIOGRAPHY PACK-LF, Medline SKU DYNJ42367B
OTHERNationwide distribution
Hazard
Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Sold states
Worldwide distribution: US (nationwide) including Puerto Rico (PR) and OUS (foreign) to countries of: Canada (CA), Netherlands (NL), Australia (AU), Korea, Republic of (KR), Sri Lanka (LK), Pakistan (PK), Japan (JP), United Arab Emirates (AE), Singapore (SG) and Slovakia (SK).
Affected count
3120 kits
Manufactured in
3 Lakes Dr, Northfield, IL, United States
Products
Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: CVOR ANGIOGRAPHY PACK-LF, Medline SKU DYNJ42367B
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1714-2026More Medline Industries, LP recalls
- FDA-PR2026-08-05Medline Hudson RCI Neonatal/Infant Heated Wire Breathing Circuits recalled for connector overheating risk
- FDA-Device2026-07-29Medline convenience kits with Bard Silastic Foley Catheters recalled for surface contamination
- FDA-Device2026-07-29Medline medical procedure kits recalled for pain medication quality issues
- FDA-Device2026-07-29Medline urology robotic procedure packs recalled for defective Bard Silastic Foley Catheters
- FDA-Device2026-07-29Medline medical convenience kits recalled for potential Bacillus contamination