FDA-Device2026-04-22Class IPROCESSING DEFECT

Insulet Omnipod 5 Pods recalled for internal cannula tear causing insulin under-delivery

OTHERNationwide distribution

Stop using affected Omnipod 5 Pods immediately

Certain Omnipod 5 Pods from 49 specific manufacturing lots contain an internal tear in the soft cannula. This defect causes insulin to leak inside the Pod instead of being delivered to your body. In some cases, the Pod may shut down completely and stop delivering insulin, or deliver less insulin than needed.

  • Check if your Pod's lot code matches any of the 49 recalled lots listed by Insulet (lot codes begin with PH1U).
  • If your Pod is from a recalled lot, stop using it immediately and switch to a backup insulin delivery method.
  • Monitor your blood glucose levels closely, especially if you notice unexplained high readings or if your Pod shuts down with a Hazard Alarm.
  • Contact Insulet or your healthcare provider for instructions on obtaining replacement Pods.
  • Seek immediate medical attention if you experience signs of high blood sugar: thirst, frequent urination, fatigue, blurry vision, nausea, or difficulty breathing.
Hazard

Due to a manufacturing defect, certain Omnipod 5 Pods from 49 lots have an internal soft cannula tear that results in insulin leaking into the Pod instead of being delivered to the user regardless of basal or bolus delivery. This defect results from damage to the unexposed portion of the soft cannula during manufacturing, which would result in a compromised fluid path. The primary failure mode is pump under-delivery due to loss of insulin to an internal leak; in some cases, the defect may also lead to pump shutoff and cessation of insulin delivery when leaked insulin contacts Pod circuity in a manner that results in an electrical short. Under-delivery of insulin (both basal and bolus insulin) or cessation of insulin put users at risk of hyperglycemia, and complications from acute and chronic hyperglycemia, including dehydration, blurry vision, nausea, vomiting, altered mental status, diabetic ketoacidosis (DKA), hyperglycemic hyperosmolar syndrome (HHS), or even death. Users may require hospitalization or medical intervention to treat severe adverse health consequences. Not all devices with the defect will issue an alarm or alert the user. If there is sufficient leakage of insulin to cause a short in the circuity, the Pod will issue a Hazard Alarm that stops all insulin delivery and alerts the user to replace their Pod. In addition, if a user s glucose is trending high and is not responding to insulin delivery, the system may reach the maximum amount of insulin microboluses allowed by the system and trigger the Automated Delivery Restriction (ADR) alert that tells users to check their blood glucose and take appropriate actions (i.e., ADR is a response to persistent hyperglycemia and maximum automated delivery constraints rather than a direct detection of the leak). The magnitude of under-delivery is unknown and based on multiple factors, including how much insulin is being delivered, whether an alarm and/or alert triggers, whether and when the user recognizes the device defect, the duration of Pod use, and the size of the tear.

Sold states
Nationwide distribution.
Affected count
1,240,115 units
Manufactured in
50 & 100 Nagog Park, N/A, Acton, MA, United States
Products
Omnipod 5 Pods. Model/Catalog number: PT-001662 ASM 5PK Pod STRL OPS G6G7

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1797-2026