FDA-Device2024-05-29Class II

Ingenia 1.5T S Model Number (REF): 781347

Philips North America Llc
Hazard

The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire

Sold states
Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Canada, China, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Italy, Japan, Lebanon, Libya, Lithuania, Martinique, Netherlands, Norway, Poland, Romania, Slovakia, Spain, Sweden, Switzerland, Thailand, United Kingdom.
Affected count
139 units
Manufactured in
222 Jacobs St, Cambridge, MA, United States
Products
Ingenia 1.5T S Model Number (REF): 781347

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1847-2024