FDA-Device2024-05-29Class II

MR 5300- Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device. Model Number (REF): (1) 782152 (2) 782110

Philips North America Llc
Hazard

The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire

Sold states
Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Canada, China, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Italy, Japan, Lebanon, Libya, Lithuania, Martinique, Netherlands, Norway, Poland, Romania, Slovakia, Spain, Sweden, Switzerland, Thailand, United Kingdom.
Affected count
81 units
Manufactured in
222 Jacobs St, Cambridge, MA, United States
Products
MR 5300- Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device. Model Number (REF): (1) 782152 (2) 782110

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1854-2024