FDA-Device2024-05-29Class II

Incisive CT - Computed Tomography X-Ray System intended to produce cross-sectional images of the body. Model Number: 728143

Philips North America Llc
Hazard

Metal mounting box on the rotating scanner on rotor (heat change box) located within the Incisive CT system, may become compromised, resulting in potential contact with other gantry components located within the Incisive CT system during rotation. Resulting in a fragment of a damaged component expelled at a low velocity

Sold states
Worldwide - US Nationwide distribution.
Affected count
18 units
Manufactured in
222 Jacobs St, Cambridge, MA, United States
Products
Incisive CT - Computed Tomography X-Ray System intended to produce cross-sectional images of the body. Model Number: 728143

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1871-2024

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief