FDA-Device2019-07-10Class II
Siemens Artis Q floor --Interventional Fluoroscopic X-Ray System Model Number: 10848280
Siemens Medical Solutions USA, Inc
Hazard
Artis Zee, Artis Q, Artis Pheno or Artis Icono systems equipped with a DCS Extended or DCS Movable, the potential exists for individual screws of the screwed connection joining the pivot bearing to the ceiling-mounted support to work loose or snap off, can result in severe injuries to patients and operating personnel
Sold states
Nationwide
Affected count
12
Manufactured in
40 Liberty Blvd, N/A, Malvern, PA, United States
Products
Siemens Artis Q floor --Interventional Fluoroscopic X-Ray System Model Number: 10848280
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1898-2019Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.