FDA-Device2019-07-31Class II
AXIOM Artis FC, Model Number 5904433
Siemens Medical Solutions USA, Inc
Hazard
A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure.
Sold states
US Nationwide Distribution
Affected count
5
Manufactured in
40 Liberty Blvd, N/A, Malvern, PA, United States
Products
AXIOM Artis FC, Model Number 5904433
Official notice