FDA-Device2021-07-28Class II

NEFA CAL (Non-Esterified Fatty Acids)- IVD Standard for measurement of NEFA Catalog Number: FA115

Randox Laboratories Ltd.
Hazard

Standard packed within NEFA kit has been assigned incorrectly. Calibrating with the incorrect value will cause a positive shift of up to +8% with both quality controls and patient samples

Sold states
US Nationwide Distribution in the states of AR, CA, CT, NY, OH, PA, TN, and TX
Affected count
71 kits US
Manufactured in
55 Diamond Road, N/A, Crumlin (North), N/A, Ireland
Products
NEFA CAL (Non-Esterified Fatty Acids)- IVD Standard for measurement of NEFA Catalog Number: FA115

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2099-2021

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief