Randox Laboratories Ltd.
24 federal recalls on file. We monitor every CPSC, FDA, USDA, and NHTSA notice for Randox Laboratories Ltd. and translate it into plain English — free weekly digest, optional same-day alerts.
- 2024-07-31FDA-DeviceClass IIRX Series CYSC- Cystatin C Reagent -Intended for in vitro diagnostic (IVD) use as Randox Cystatin C assay on automated analysers. Catalogue Number: CYS4004
Cystatin C Reagent marketed without a 510 (k)
- 2024-05-15FDA-DeviceClass IIIEvidence MultiSTAT with software v 3.7-Analyzer intended for the qualitative determination of parent drug molecule and metabolites of drugs in human urine Catalog Number: EV4115
Foam gasket on the Evidence MultiSTAT chip heater assembly may not perform as expected. Foam that is not functioning as expected may cause light to leak into the Charged Coupled Device (CCD) Camera during imaging. This will affect the testing and may delay in reporting results.
- 2024-05-15FDA-DeviceClass IIClinical Chemistry Calibration Serum (Catalog number Cal2351) that is 20x5ml lyophilized human based serum covering 38 commonly used clinical chemistry tests.
Randox Laboratories has identified that Inorganic Phosphate in Calibration Serum Level 3, CAL2351 & CAL10388, is running with a negative bias on RX Series instruments compared to other methods.
- 2024-04-24FDA-DeviceClass IIMicroalbumin Calibrator Series (mALB CAL)-IVD to calibrate the Microalbumin Assay Catalog Number: MA2426
Microalbumin Calibrator Series, MA1567 & MA2426, is running with a negative bias compared to other method may misclassify patient results. Users may experience a shift in Quality Control and patient sample recovery of up to +12%.
- 2024-04-24FDA-DeviceClass IIMicroalbumin Calibrator Series (mALB CAL)-IVD used to calibrate the Microalbumin Assay Catalog Number: MA1567
Microalbumin Calibrator Series, MA1567 & MA2426, is running with a negative bias compared to other methods and may misclassify patient results. Users may experience a shift in Quality Control and patient sample recovery of up to +12%.
- 2024-04-03FDA-DeviceClass IICystatin C Calibrator - In vitro diagnostic (IVD) use in the calibration of the Randox Cystatin C assay on automated analysers Catalog Number: CVS 2699
Cystatin C Calibrator Series CVS 2699, is running with a negative bias compared to other methods
- 2024-03-27FDA-DeviceClass IILiquid Protein Calibrators. C3 and Haptoglobin liquid protein calibrators used with Randox IgA, IgG, and IgM assays that require sample predilution.
The firm has realigned C3 and Haptoglobin in Liquid Protein Calibrators, IT2691, to reference material ERM-DA470k/IFCC.
- 2024-02-14FDA-DeviceClass IIRX Series Copper (Cu) Assay Ref. Number CU2340
Randox Laboratories have released an update to the Carryover Avoidance Technical Bulletin to detail that Copper, should not be run directly after Total Protein, on the RX series instruments testing order. If copper assay is run after total protein it could lead to falsely elevated copper results.
- 2024-01-24FDA-DeviceClass IICalibration Serum Level 3 - IVD For use as a Calibrator in clinical chemistry assays Catalog Number: CAL2351
Target for Alkaline Phosphatase (ALP) for the AMP optimized to IFCC 37¿C method has been mis-assigned for the RX Series instruments by approximately 10% on Quality Control and patient samples
- 2024-01-24FDA-DeviceClass IICalibration Serum Level 3 CAL2351
CK Total in Calibration Serum Level 3, CAL2351, is running with a positive bias on RX Series instruments compared to other methods experience a shift in Quality Control and patient sample recovery of up to 13%.
- 2024-01-24FDA-DeviceClass IICalibration Serum Level 3. CAL 3, REF CAL2351. Human Assayed Multi-Sera Level 2 (HN1530) Human Assayed Multi-Sera Level 3 (HE 1532)
There have been transcription errors on the Instructions For Use (IFU) of the Calibration Serum Level 3 (CAL2351 lot 1260UE, 1262UE and 1315UE) for target values of Bilirubin and Triglycerides, on the IFU of Human Assayed Multi-Sera Level 2 (HN1530) for TBIC taget value and on IFU of Human Assayed Multi-Sera Level 3 (HE 1532) for Lipase target value.
- 2024-01-10FDA-DeviceClass IIISEROLOGY ToRCH lgM 2022 POSITIVE CONTROLI Intended for use with in vitro assays for determination of IgM antibodies to Cytomegalovirus (CMV), IgM antibodies to Rubella Virus, IgM antibodies to Toxoplasma gondii, IgM antibodies to Herpes Simplex Virus Type 1 and 2. Catalog Number: SR10349
A decrease in the reactivity of HSV Type 1/2 lgM in the Serology ToRCH lgM Positive Control when tested on the DiaSorin Liaison XL.
- 2023-12-13FDA-DeviceClass IIRANDOX Total Bilirubin.
Randox has had reports of elevated patient results using Plasma (lithium heparin) samples with Total Bilirubin reagent (TBil) Vanadate Oxidation Method.
- 2023-11-08FDA-DeviceClass IIRX Monaco-IVD Clinical Chemistry Analyser Catalog Number: RX5000
RX Parameter rerun settings (Sample Volume Low/High), and Dilutions used for Auto reruns, not meeting sample re-run conditions when samples are above or below assay range and may result i issues with the re-run feature conditions when incorrectly. Potential to report incorrectly.
- 2023-11-08FDA-DeviceClass IIRX Modena-IVD Clinical Chemistry Analyser Catalog Number: RX9000
RX Parameter rerun settings (Sample Volume Low/High), and Dilutions used for Auto reruns, not meeting sample re-run conditions when samples are above or below assay range and may result i issues with the re-run feature conditions when incorrectly. Potential to report incorrectly.
- 2023-11-08FDA-DeviceClass IIRX Imola-IVD Clinical Chemistry Analyzer Catalog Number: RX4900. RX4900R (refurbished)
RX Parameter rerun settings (Sample Volume Low/High), and Dilutions used for Auto reruns, not meeting sample re-run conditions when samples are above or below assay range and may result i issues with the re-run feature conditions when incorrectly. Potential to report incorrectly.
- 2023-11-08FDA-DeviceClass IIRX Daytona Plus (without ISE)- IVD Clinical Chemistry Analyser (1) RX4041, (2) RX4041R (refurbished)
RX Parameter rerun settings (Sample Volume Low/High), and Dilutions used for Auto reruns, not meeting sample re-run conditions when samples are above or below assay range and may result i issues with the re-run feature conditions when incorrectly. Potential to report incorrectly.
- 2023-11-08FDA-DeviceClass IIRX Daytona Plus (with ISE)-IVD Clinical Chemistry Analyser Catalog Number: RX4040
RX Parameter rerun settings (Sample Volume Low/High), and Dilutions used for Auto reruns, not meeting sample re-run conditions when samples are above or below assay range and may result i issues with the re-run feature conditions when incorrectly. Potential to report incorrectly.
- 2023-09-27FDA-DeviceClass IILiquid Assayed Chemistry Control Premium Plus Level 3, Catalog Number LAE4215
there has been a transcription error in the Instructions for Use (IFU) for the Liquid Assayed Chemistry Control Premium Plus Level 3. For Caeruloplasmin, the one and two standard deviation values printed on the sheet are incorrect, which may lead to misreporting patient results. Target and range values are correct, and risk is mitigated; most clinical labs will use the target and range values in clinical practice.
- 2023-08-30FDA-DeviceClass IIRandox Liquid Protein Calibrators-SP CAL-(LIQ)-In vitro diagnostic product used for the calibration of ASO, Complement C3, Complement C4, CRP, Ferritin, Haptoglobin, IgM, Prealbumin, RF and Transferrin assays Catalogue Number: IT2691
As part of the firm's ongoing quality monitoring (see internal complaint # OCC72574), Randox Laboratories have restandardized Ferritin in Liquid Protein Calibrators, IT2691, to reference material NISBC 19/118. The calibrators lots 2112IT-2116IT, packed into batches 627222, 627224, 634886 and 634887 have been reassigned as part of the restandardization. Following this restandardization, Ferritin results for Quality Control material and patient samples recovered erroneously higher than the targeted calibrator values by approximately +10% across the assay range following.
- 2023-08-30FDA-DeviceClass IIRandox Liquid Urine Control Level 2- intended for in vitro diagnostic use, in the quality control of Amylase, Calcium, Chloride, Cortisol, Creatinine, Glucose, hCG pregnancy, Magnesium, Microalbumin, Osmolality, pH, Phosphate Inorganic, Potassium, Protein Total, Sodium, Specific Gravity, Urea and Uric Acid (Urate) on clinical chemistry systems Catalog Number: UC5074
(1) There is vial to vial variation resulting in some vials recovering positive for hCG, which should be negative and (2) high and outside range for cortisol. (3) There has been a transcription error for Creatinine in the Instructions For Use (IFU). The target and ranges for Creatinine for the Roche Creatinine Plus method have been listed incorrectly. These failures cause a delay in patient results for the above mentioned analytes.
- 2023-08-16FDA-DeviceClass IIRandox RX Series Urea Enzymatic Kinetic Assay Ref. Numbers: (1) UR3825 (2) UR3873 (3) UR8334 (4) UR8070 (Modena)
Carryover from the Direct LDL-cholesterol assay on the Rx instruments may result in a negative bias, up to -11%, in quality control and patient samples, in urea assays results if the urea assay is run immediately after the Direct LDL-cholesterol assay. This may lead to a delay in reporting results or in falsely decreased urea test results.
- 2023-07-05FDA-DeviceClass IIIImmunoassay Premium Control Tri Level-In vitro diagnostic use, as assayed quality control material to monitor the accuracy and reproducibility of analytes Catalog Number: IA2633
Delay in reporting results due to Quality Controls running high outside of range. Target value and range for ACTH in Immunoassay Premium Level 3, IA2640, lot 2031EC and Immunoassay Premium Control Tri Level kit batch 583135 lot 2031EC has been reassigned on the Roche Cobas e801 due to High recovery outside range
- 2023-07-05FDA-DeviceClass IIIImmunoassay Premium Level 3-In vitro diagnostic use, as assayed quality control material to monitor the accuracy and reproducibility of analytes. Catalog Number: IA2640
Delay in reporting results due to Quality Controls running high outside of range. Target value and range for ACTH in Immunoassay Premium Level 3, IA2640, lot 2031EC and Immunoassay Premium Control Tri Level kit batch 583135 lot 2031EC has been reassigned on the Roche Cobas e801 due to High recovery outside range
Get notified about new Randox Laboratories Ltd. recalls
Free weekly digest. We email when a new federal recall affects Randox Laboratories Ltd. or anything else in your household.
Get the Sunday Brief