FDA-Device2020-06-03Class II
Cios Alpha Image-Intensified Fluoroscopic X-Ray System, Model Numbers VA20-10308191 and VA30-10308194 The Cios Alpha and Cios Spin are a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications.
Siemens Medical Solutions USA, Inc
Hazard
The unplugging of the main cable on the monitor trolley from the X10 connector can cause an unsafe electrical voltage (up to 67,6 V DC) at the X10 connector. This can result in an electrical shock if the user or other persons touch the contact pins of the X10 connector.
Sold states
US Nationwide Distribution to states of: AR, AZ, CA, FL, IL, IN, MA, MD, MI, MN, NC, NE, NH, NJ, NY, OH, OK, SC, TN, TX, UT, VA, WA, and Puerto Rico.
Affected count
58
Manufactured in
40 Liberty Blvd, N/A, Malvern, PA, United States
Products
Cios Alpha Image-Intensified Fluoroscopic X-Ray System, Model Numbers VA20-10308191 and VA30-10308194 The Cios Alpha and Cios Spin are a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2127-2020Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief