FDA-Device2024-07-10Class II
Zyno Medical Z800, Z800F, Z800W, and Z800WF Infusion Pumps
Zyno Medical LLC
Hazard
Zyno Medical has found that certain Z800, Z800F, Z800W, and Z800WF pumps have faulty power supply modules that will not provide a charge to the rechargeable battery within the pump. This may lead to the battery having insufficient or no charge available to the pump when it is unplugged from the AC mains power.
Sold states
US Nationwide distribution.
Affected count
88 units
Manufactured in
177 Pine St, N/A, Natick, MA, United States
Products
Zyno Medical Z800, Z800F, Z800W, and Z800WF Infusion Pumps
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2263-2024Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief