FDA-Device2025-08-27Class I
RELIANCE 4-FRONT, endocardial cardioversion/defibrillation and pace/sense lead, Permanent Defibrillator Electrodes; Model Nos. 0654, 0655, 0657, 0658, 0682, 0683, 0685, 0686, 0692, 0693, 0695, 0696
Hazard
Affected lead models may experience gradually rising LVSI measurements in association with calcification of the defibrillation lead coils. The issue may lead to early lead replacement. Most serious harm is death or need for cardiac resuscitation.
Sold states
Worldwide Distribution. US nationwide. International distribution worldwide.
Affected count
247,702 US; 344,670 OUS
Manufactured in
4100 Hamline Ave N, N/A, Saint Paul, MN, United States
Products
RELIANCE 4-FRONT, endocardial cardioversion/defibrillation and pace/sense lead, Permanent Defibrillator Electrodes; Model Nos. 0654, 0655, 0657, 0658, 0682, 0683, 0685, 0686, 0692, 0693, 0695, 0696
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2327-2025More Boston Scientific Corporation recalls
- FDA-Device2026-07-29Boston Scientific FARADRIVE Steerable Sheath recalled for air embolism risk from valve damage
- FDA-Device2026-07-29Boston Scientific FARADRIVE Steerable Sheath recalled for air embolism risk from valve damage
- FDA-Device2026-07-29Boston Scientific FARADRIVE Steerable Sheath recalled for air embolism risk from valve damage
- FDA-Device2026-07-29Boston Scientific TheraSphere 360 Y-90 software anomaly may cause incorrect dose calculations
- FDA-Device2026-07-29Boston Scientific FARASTAR PFA Generator recalled for incomplete labeling
Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to your household — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief