Boston Scientific TheraSphere 360 Y-90 software anomaly may cause incorrect dose calculations
The TheraSphere 360 Y-90 Management Platform has a software bug that can cause incorrect dose calculations. When a user modifies calculation inputs and recalculates, the new dose is computed correctly on screen, but vials previously added to the cart do not automatically update to match the revised calculation.
From the FDA notice · Read the official notice ↗ · All Boston Scientific recalls
What to do
- Contact Boston Scientific for guidance on affected software versions V1.1.2, V1.2.1, V1.2.2, V2.0.1, and V2.0.2
- Verify that vials in the cart match the most recent calculation before use
- Report any calculations or dosing errors to Boston Scientific immediately
Is yours affected?
Compare your label or packaging with the recalled product.
- Product
- Boston Scientific TheraSphere 360 Y-90 Management Platform, Model Number M00139509001; Radiological Image Processing System.
- Sold in
- Nationwide
- Made for
- All ages (0 months and older)
Details
- Units affected
- 68 units
- Illnesses reported
- None reported
- Manufactured in
- 1 Scimed Pl, N/A, Maple Grove, MN, United States
- Classification
- Class II
- Recall number
- Z-2771-2026
In the FDA's words
- Description
- Boston Scientific TheraSphere 360 Y-90 Management Platform, Model Number M00139509001; Radiological Image Processing System. Software anomaly occurs when a user adds one or more vials to the cart and subsequently returns to the Activity Calculator to modify one or more input. When a user modifies an input and selects "Recalculate", the dose calculation is updated correctly; however, any vials that were previously added to the cart are not automatically updated to reflect the revised calculation. UDI/DI 00191506068264, All released software versions (V1.1.2, V1.2.1, V1.2.2. V2.0.1, V2.0.2) Distribution: US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, FL, GA, IL, IN, LA, MA, MN, NE, NY, OH, OK, OR, SC, SD, TN, TX, UT, WA, WI, WV, WY. Affected quantity: 68 units
- Hazard
- Software anomaly occurs when a user adds one or more vials to the cart and subsequently returns to the Activity Calculator to modify one or more input. When a user modifies an input and selects "Recalculate", the dose calculation is updated correctly; however, any vials that were previously added to the cart are not automatically updated to reflect the revised calculation.
More from Boston Scientific
See all- Boston Scientific Imager II catheters recalled for tip fracture risk
SeriousFDASep 30, 2026
- Boston Scientific ENROUTE system recalled over sheath tip separation
SeriousFDAAug 26, 2026