FDA-Device2026-07-29Class IIPROCESSING DEFECT

Boston Scientific TheraSphere 360 Y-90 software anomaly may cause incorrect dose calculations

OTHERNationwide distribution

Boston Scientific TheraSphere 360 Y-90 dose calculation software issue

The TheraSphere 360 Y-90 Management Platform has a software bug that can cause incorrect dose calculations. When a user modifies calculation inputs and recalculates, the new dose is computed correctly on screen, but vials previously added to the cart do not automatically update to match the revised calculation.

  • Contact Boston Scientific for guidance on affected software versions V1.1.2, V1.2.1, V1.2.2, V2.0.1, and V2.0.2
  • Verify that vials in the cart match the most recent calculation before use
  • Report any calculations or dosing errors to Boston Scientific immediately
Hazard

Software anomaly occurs when a user adds one or more vials to the cart and subsequently returns to the Activity Calculator to modify one or more input. When a user modifies an input and selects "Recalculate", the dose calculation is updated correctly; however, any vials that were previously added to the cart are not automatically updated to reflect the revised calculation.

Sold states
US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, FL, GA, IL, IN, LA, MA, MN, NE, NY, OH, OK, OR, SC, SD, TN, TX, UT, WA, WI, WV, WY.
Affected count
68 units
Manufactured in
1 Scimed Pl, N/A, Maple Grove, MN, United States
Products
Boston Scientific TheraSphere 360 Y-90 Management Platform, Model Number M00139509001; Radiological Image Processing System.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2771-2026

More Boston Scientific Corporation recalls