FDA-Device2026-06-17Class IIPROCESSING DEFECT

Elekta MOSAIQ Oncology Information System software may fail to load treatment plans during communication errors

OTHERNationwide distribution

Elekta MOSAIQ cancer treatment software may fail to load plans

Elekta's MOSAIQ Oncology Information System software (version 3.2) may fail to load an adapted baseline shift plan during cancer treatment delivery if a communication error occurs. This could affect the accuracy of radiation therapy positioning. No injuries have been reported.

  • If you operate or manage this system, contact Elekta immediately for guidance on this issue
  • Review your institution's communication protocols and error-handling procedures for treatment delivery
  • Monitor treatment sessions for any signs of plan-loading errors
Hazard

Software may fail to load adapted baseline shift plan during treatment delivery when a communication error occurs.

Sold states
Worldwide - US Nationwide distribution in the states of WI, FL, LA, NY, NY, NJ, KS, FL, IA, NM, TX and the countries of Bahrain, Belgium, Canada, China, Denmark, France, Germany, Hong Kong, India, Israel, Italy, Japan, Lithuania, Malta, Mexico, Netherlands, Pakistan, Panama, Poland, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Kingdom.
Affected count
12 programs
Manufactured in
400 Perimeter Center Ter NE Ste 50, N/A, Atlanta, GA, United States
Products
MOSAIQ Oncology Information System

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2406-2026

More Elekta, Inc. recalls