FDA-Device2018-07-25Class II
t:slim G4 Insulin Pump with Dexcom G4 Platinum CGM Software version: 4.3.4.3, Firmware version: 004722, Firmware, Fuel Gauge, Binary Rev B.
Tandem Diabetes Care Inc
Hazard
The fuel gauge, the component that reads and reports the battery parameters to the pump, could provide inaccurate readings, which present to the user in one of two ways: A) by triggering a Malfunction 4 Alarm, or B) by triggering a succession of notifications prior to the pump shutting off, including Alert 2 (Battery Low), Alert 3 (Battery Very Low) and Alarm 12 (Battery Very Low). The alarms in both scenarios notify the user that the pump has stopped delivering insulin.
Sold states
U.S. Consignees: IN, AZ, UT, TX, NY, NJ, MD, MN, IL, CO, AL, WI, CA, PA, OH, OR, MS, FL, NM, ID, NV, KY, VA,
Affected count
55
Manufactured in
11075 Roselle St Ste 200, San Diego, CA, United States
Products
t:slim G4 Insulin Pump with Dexcom G4 Platinum CGM Software version: 4.3.4.3, Firmware version: 004722, Firmware, Fuel Gauge, Binary Rev B.