- 2025-10-29FDA-DeviceClass IIt:slim X2 insulin pump with Interoperable Technology, Control-IQ+ technology, REF: 1014070, 1014071, 1014072, 1014073, and User Guide when using Dexcom G7 sensor; Tandem Mobi insulin pump with Interoperable Technology, Control-IQ+ technology, REF: 1014801, 1014811, 1014813, and User Guide when using Dexcom G7 sensor
A software defect in versions 7.9.0.1 and 7.10.1 of the pump software for Tandem t:slim X2 and version 7.9.0.1 of the Tandem Mobi pump with Control-IQ+ technology that are paired with a G7 sensor may result in an unexpected automatic insulin correction bolus (autobolus ) which could result in hypoglycemia.
- 2025-10-01FDA-DeviceClass IIt:slim X2 Insulin Pump with Interoperable Technology (with Tandem t:slim mobile app)
an app defect that occurs when the phone is set to a right-to-left language, which causes app-pump pairing issues and graphical defects, which can potentially lead to incorrect therapy decisions and subsequent hypoglycemia or hyperglycemia.
- 2025-09-03FDA-DeviceClass IBattery Operated infusion pumps. t:slim X2 G5, Replacement, Refurbished, REF: 1000911; t:slim X2 GS Classic, Replacement, Refurbished, REF: 1003808; t: slim X2, Basal-IQ, mg/dl, REF: 1004219; Pump, t:slim X2, Clinical Use Only, REF: 1004484; Pump, t:slim X2, Control-IQ, REF: 1005611; Pump, t:slim X2, Control-IQ, mmoUL, Refurbished Replacement, REF: 1005613; Pump, t:slim X2, Control-IQ, mg/dl, REF: 1005615; Pump, t:slim X2, Control-IQ, mg/dl, Refurbished Replacement, REF: 1005617; Pump,
Insulin pumps manufactured with speakers from Revision A and B may malfunction ("Malfunction 16") which prevents audible alerts and causes the pumps to become inoperable and stop infusing insulin which could result in a hyperglycemic event, which may lead to injury and/or hospitalization.
- 2025-04-23FDA-DeviceClass ITandem Mobi Insulin Pump with Interoperable Technology
A software defect in Version 7.9 of the pump software for Tandem t:slim X2 and Tandem Mobi pumps, when used with Control IQ+ technology, will cause the pump to incorrectly interpolate glucose trends when the Estimated Glucose Value (EGV) is above 255 mg/dL at the start or end of a gap in data collection due to a lapse in connection from a paired continuous glucose monitor (CGM) sensor, which can lead to under-delivery or over-delivery of insulin based on inaccurate result leading to severe cases of hypoglycemia or hyperglycemia.
- 2025-04-23FDA-DeviceClass It:slim X2 Insulin Pump with Interoperable Technology
A software defect in Version 7.9 of the pump software for Tandem t:slim X2 and Tandem Mobi pumps, when used with Control IQ+ technology, will cause the pump to incorrectly interpolate glucose trends when the Estimated Glucose Value (EGV) is above 255 mg/dL at the start or end of a gap in data collection due to a lapse in connection from a paired continuous glucose monitor (CGM) sensor, which can lead to under-delivery or over-delivery of insulin based on inaccurate result leading to severe cases of hypoglycemia or hyperglycemia.
- 2024-05-15FDA-DeviceClass It:connect mobile app used in conjunction with t:slim X2 insulin pump with Control-IQ technology
During normal use, the mobile app version 2.7 may crash and be automatically relaunched by the iOS operating system. This cycle intermittently repeats, which leads to excessive Bluetooth communication that may result in pump battery drain and may lead to the pump shutting down sooner than typically expected. Pump shutdown will cause insulin delivery to suspend, which could lead to an under-delivery of insulin and may result in hyperglycemia, including severe hyperglycemia.