FDA-Device2021-09-22Class II
The Azurion series (within the limits of the used Operation Room table) are intended for use to perform: "Image guidance in diagnostic, interventional and minimally invasive surgery procedures for the following clinical application areas: vascular, non-vascular, cardiovascular and neuro procedures. "Cardiac imaging applications including diagnostics, interventional and minimally invasive surgery procedures. Additionally: "The Azurion series can be used in a hybrid Operation Room. "The Azuri
Hazard
Early life failure of the ADUs only occurs when the hospital mains provide 480V to the system. This current voltage may lead to saturation and overheating of the coils that protect the IGBTs (insulated-gate bipolar transistor) of the ADU. The overheating may generate a peculiar burning smell that may be noticed. In some instances, this burning smell may be noticed before the ADU fails. When the ADU fails the X-ray performance of the system is reduced to Emergency Fluoroscopy. Exposure will not be possible and image quality will be reduced.
Sold states
US Nationwide distribution in the states of ID, MA, MI, TX, VA.
Affected count
1 unit
Manufactured in
222 Jacobs St, N/A, Cambridge, MA, United States
Products
The Azurion series (within the limits of the used Operation Room table) are intended for use to perform: "Image guidance in diagnostic, interventional and minimally invasive surgery procedures for the following clinical application areas: vascular, non-vascular, cardiovascular and neuro procedures. "Cardiac imaging applications including diagnostics, interventional and minimally invasive surgery procedures. Additionally: "The Azurion series can be used in a hybrid Operation Room. "The Azurion series contain a number of features to support a flexible and patient centric procedural workflow.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2476-2021More Philips North America Llc recalls
- FDA-Device2026-07-01Philips Avalon FM20 Fetal Monitor recalled for speaker connector assembly defect
- FDA-Device2026-07-01Philips Avalon Fetal Monitor, FM 30 Part numberM2703A
- FDA-Device2026-07-01Philips Telemetry Monitor 5500 1.4 GHz. Model Number: 867232. Includes the following system descriptions: TELE Monitor 5500, 1.4 Ghz, ECG only, ex; Telemetry Monitor 5500 1.4 GHz FAST EX.
- FDA-Device2026-06-17Philips DigitalDiagnost C90 Flex/Value/Chest/ER recalled for missing seismic stabilization kit
Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief