Medline Nursing Kit (EDUC1044) recalled for catheter defects and connector detachment risk
Medline Nursing Kit recalled for catheter defects
Medline identified quality issues in certain Nursing Kits (SKU EDUC1044) during an internal review. The catheter components did not meet specifications—including surface irregularities, connectors that detached prematurely, and some catheters that were the wrong diameter. About 730 kits were affected.
- Check if you have a Medline Nursing Kit with SKU EDUC1044, UDI 10195327449476 or 40195327449477, and Lot 23HDA449.
- Contact Medline or your supplier for replacement or disposal instructions.
- Do not use the kit if you have one, as connector or catheter failure could compromise patient care.
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2499-2026More Medline recalls
- FDA-PR2026-08-05Medline Hudson RCI Neonatal/Infant Heated Wire Breathing Circuits recalled for connector overheating risk
- FDA-Device2026-07-29Medline surgical convenience kits recalled for quality issues in anesthetic component
- FDA-Device2026-07-29Medline NEURO PACK medical kits recalled for Bacillus contamination
- FDA-Device2026-07-29Medline convenience kits with Bard Silastic Foley Catheters recalled for surface contamination