Moderate riskFDA DeviceClass II

Inspire Respiratory Sensing Lead Model 4340 recalled for incorrect expiration dating

Inspire Medical Systems' Respiratory Sensing Lead (Model 4340) was labeled with expiration dates that extend beyond the validated three-year shelf life. If the device is used after its actual expiration date, the material may degrade and could require unplanned revision surgery.

From the FDA notice · Read the official notice ↗

What to do

  1. Check your implanted device paperwork for the serial number and Use By date
  2. Contact your surgeon if you received this device with a serial number listed in the recall
  3. Do not use any unit past its three-year shelf life from the manufacturer date

Is yours affected?

Compare your label or packaging with the recalled product.

Product
Brand Name: Inspire Medical Systems, Inc. Product Name: Inspire Respiratory Sensing Lead Model/Catalog Number: 4340 Software Version: N/A Product Description: The Inspire¿ Respiratory Sensing Lead (Model 4340) is designed to detect respiratory effort. The lead features a pressure sensitive membrane that converts the mechanical energy of respiration into an electrical signal. The lead incorporates a standard connector for coupling to the implantable pulse generator (IPG). Component: N/A
Sold in
Nationwide
Made for
Adults (18 years to 120 years)

This recall covers 2 products — see all of them.

Details

Units affected
50 impacted units
Illnesses reported
None reported
Manufactured in
5500 Wayzata Blvd Ste 1600, Golden Valley, MN, United States
Classification
Class II
Recall number
Z-2515-2026

In the FDA's words

Description
Brand Name: Inspire Medical Systems, Inc. Product Name: Inspire Respiratory Sensing Lead Model/Catalog Number: 4340 Software Version: N/A Product Description: The Inspire¿ Respiratory Sensing Lead (Model 4340) is designed to detect respiratory effort. The lead features a pressure sensitive membrane that converts the mechanical energy of respiration into an electrical signal. The lead incorporates a standard connector for coupling to the implantable pulse generator (IPG). Component: N/A Products are labeled with incorrect Use By Date, extending beyond the validated three-year shelf life from date of manufacturer. Material degradation of a product past its Use By Date may result in unplanned or premature revision surgery. Lot Code: Model No 4340 (USA only lead); UDI-DI [4340] 10855728005946; Inspire part number [4340] 900-014-003; serial numbers T18806, T45012, T45017, T49002, T67098, T67152, T69541, T69542, T69553, T69558, T69562 Model No 4340 (Global lead); UDI-DI [4340] 10810098650301; Inspire part number [4340] 900-014-005; serial numbers T91485, T91486, T91491, T91495, T91513, T91520, T91521, T91527, T91533, T91537, T91540, T91545, T96165, T96168, T96192, T96193, T96195, T96225, T96237, T98118, T98126, T98136, T98137, T98138, T98141, T98143, T98145, T98150, T98153, T98156, T98163, T98165, T98176, T98177, T98178, T98181, T98183, T98190, T96217 Distribution: Worldwide distribution - US Nationwide and the country of Hong Kong. Affected quantity: 50 impacted units
Hazard
Products are labeled with incorrect Use By Date, extending beyond the validated three-year shelf life from date of manufacturer. Material degradation of a product past its Use By Date may result in unplanned or premature revision surgery.

2 products in this recall

Inspire respiratory and stimulation leads recalled for mislabeled Use By Date

Inspire Medical Systems is recalling two implant lead products, the Respiratory Sensing Lead and the Stimulation Lead, because they carry a Use By Date that goes beyond their validated three-year shelf life. The agency says material can degrade past the labeled date, which may lead to unplanned or premature revision surgery. No injuries have been reported.

The agency lists each product separately. Check every one you might have.

Related topics

Expiration date errorMedical devicesSold nationwideMore from FDA Device