Ventec Life Systems VOCSN+Pro ventilator recalled for faulty backup battery alarm
Ventec Life Systems' VOCSN+Pro respiratory ventilator may have a damaged backup battery that powers the alarm system. If primary power is lost and the backup battery is depleted or absent, the backup alarm may not sound to alert caregivers of the power loss.
From the FDA notice · Read the official notice ↗ · All Ventec Life Systems recalls
What to do
- Check your VOCSN+Pro device serial number against the lot number 804613 or 807138 listed in the recall notice.
- Verify the backup battery is functioning and fully charged before each use.
- Contact Ventec Life Systems immediately if your unit is affected to arrange inspection or replacement.
- Ensure a caregiver is present during operation and do not rely solely on the alarm system.
Is yours affected?
Compare your label or packaging with the recalled product.
- Product
- VOCSN+Pro package, REF: PRT-00853-000; VOCSN+Pro unit, REF: PRT-00490-001
- Sold in
- Nationwide
- Made for
- Adults (0 months and older)
Details
- Units affected
- 1
- Illnesses reported
- None reported
- Manufactured in
- 22002 26th Ave Se, N/A, Bothell, WA, United States
- Classification
- Class II
- Recall number
- Z-2525-2026
In the FDA's words
- Description
- VOCSN+Pro package, REF: PRT-00853-000; VOCSN+Pro unit, REF: PRT-00490-001 Respiratory system intended to provide continuous/intermittent ventilator support, may have a damaged back-up battery supporting backup alarm function, which may result in the backup alarm not activating during loss of primary power and absence/depletion of backup power sources, which may result in delayed caregiver awareness of a hazardous condition. REF/UDI-DI/Serial Number/Lot: PRT-00853-000/00855573007839/155391P/807138, PRT-00490-001/00855573007792/155391/804613 Distribution: US Nationwide distribution in the state of UT. Affected quantity: 1
- Hazard
- Respiratory system intended to provide continuous/intermittent ventilator support, may have a damaged back-up battery supporting backup alarm function, which may result in the backup alarm not activating during loss of primary power and absence/depletion of backup power sources, which may result in delayed caregiver awareness of a hazardous condition.