Serious riskFDA DeviceClass II

Ventec Life Systems VOCSN+Pro ventilator recalled for faulty backup battery alarm

Ventec Life Systems' VOCSN+Pro respiratory ventilator may have a damaged backup battery that powers the alarm system. If primary power is lost and the backup battery is depleted or absent, the backup alarm may not sound to alert caregivers of the power loss.

From the FDA notice · Read the official notice ↗ · All Ventec Life Systems recalls

What to do

  1. Check your VOCSN+Pro device serial number against the lot number 804613 or 807138 listed in the recall notice.
  2. Verify the backup battery is functioning and fully charged before each use.
  3. Contact Ventec Life Systems immediately if your unit is affected to arrange inspection or replacement.
  4. Ensure a caregiver is present during operation and do not rely solely on the alarm system.

Is yours affected?

Compare your label or packaging with the recalled product.

Product
VOCSN+Pro package, REF: PRT-00853-000; VOCSN+Pro unit, REF: PRT-00490-001
Sold in
Nationwide
Made for
Adults (0 months and older)

Details

Units affected
1
Illnesses reported
None reported
Manufactured in
22002 26th Ave Se, N/A, Bothell, WA, United States
Classification
Class II
Recall number
Z-2525-2026

In the FDA's words

Description
VOCSN+Pro package, REF: PRT-00853-000; VOCSN+Pro unit, REF: PRT-00490-001 Respiratory system intended to provide continuous/intermittent ventilator support, may have a damaged back-up battery supporting backup alarm function, which may result in the backup alarm not activating during loss of primary power and absence/depletion of backup power sources, which may result in delayed caregiver awareness of a hazardous condition. REF/UDI-DI/Serial Number/Lot: PRT-00853-000/00855573007839/155391P/807138, PRT-00490-001/00855573007792/155391/804613 Distribution: US Nationwide distribution in the state of UT. Affected quantity: 1
Hazard
Respiratory system intended to provide continuous/intermittent ventilator support, may have a damaged back-up battery supporting backup alarm function, which may result in the backup alarm not activating during loss of primary power and absence/depletion of backup power sources, which may result in delayed caregiver awareness of a hazardous condition.