FDA-Device2021-10-06Class II

Covidien HysteroLux Fluid Management System Control Unit, REF: 72205000

WOM World of Medicine AG
Hazard

When the display of inflow volume to the uterus reaches 32450 ml, the fluid management system control unit display will freeze because the internal software calculation threshold is reached. The outflow measurement will continue, and so the deficit accumulated up to this point will start counting backwards to 0 ml. There is a risk of distention fluid reaching the soft tissue circulatory system.

Sold states
US Nationwide distribution in the state of CT.
Affected count
1100
Manufactured in
Salzufer 8, N/A, Berlin, N/A, Germany
Products
Covidien HysteroLux Fluid Management System Control Unit, REF: 72205000

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2561-2021

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief