Brand
WOM World Of Medicine AG
2 federal recalls on file. We monitor every CPSC, FDA, USDA, and NHTSA notice for WOM World Of Medicine AG and translate it into plain English — free weekly digest, optional same-day alerts.
- 2024-07-10FDA-DeviceClass IIWOM Tube Set for lrrlgatlon, single-use- Intended to transport irrigation fluid to the patient in endoscopic surgeries. Distr. by Endoscopy Smith & Nephew Inc. REF Numbers: T0449-01
Incorrect transponder data was written on the RFID tags triggers the error message E03, Tube Set lnvalid, may result in a delay of surgery
- 2023-10-04FDA-DeviceClass IIAquilex Fluid Control System-Intended to provide fluid distension of the uterus during diagnostic and operative hysteroscopies and to monitor the volume differential between the irrigation fluid flowing into and out of the uterus. Models AQL-100PBS and AQL-100P
The display of inflow volume can reach its limit of 30,000 ml during long procdeure and the inflow volume display will freeze at the maximum value while the deficit will start counting backwards until 0 ml is reached and may result in fluid overload