Dexcom G7 glucose sensors recalled for infection and malfunction risks
Certain lots of Dexcom G7 continuous glucose monitoring sensors that were designated for destruction were stolen and later sold to pharmacies and distributors without authorization. Two affected lots have different risks: one was not properly sterilized and poses an infection risk, and the other may not provide accurate readings or detect dangerous blood sugar levels.
From the FDA notice · Read the official notice ↗
What to do
- Check your Dexcom G7 sensor packaging for lot number 1725204004 or 1725069002
- If you have either lot, stop using the sensor and contact Dexcom or your healthcare provider
- Do not discard the sensor; keep it to return to your pharmacy or Dexcom
Is yours affected?
Compare your label or packaging with the recalled product.
- Product
- Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 US Retail Sensor/Transmitter 1-pack Model/Catalog Number: STP-AT-012 Software Version: N/A Product Description: Dexcom G7 US Retail Sensor/Transmitter 1-pack Component: Dexcom G7 Continuous Glucose Monitoring System
- Sold in
- Nationwide
- Made for
- Adults (0 months and older)
Details
- Units affected
- N/A
- Illnesses reported
- None reported
- Manufactured in
- 6340 Sequence Dr, N/A, San Diego, CA, United States
- Classification
- Class II
- Recall number
- Z-2574-2026
In the FDA's words
- Description
- Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 US Retail Sensor/Transmitter 1-pack Model/Catalog Number: STP-AT-012 Software Version: N/A Product Description: Dexcom G7 US Retail Sensor/Transmitter 1-pack Component: Dexcom G7 Continuous Glucose Monitoring System Certain lots of Dexcom G7 sensors originally designated as scrap and intended for destruction were stolen during the destruction process. The sensors were later sold to pharmacies and distributors by Pharmsource, LLC. Lot 1725204004 were designated for destruction because the G7 sensors were not properly sterilized and have an increased risk of skin infection if used. Lot 1725069002 were designated for destruction due to an increased risk of having no sensor readings available or experiencing sensor inaccuracy that includes the potential for missed detection of a hyperglycemic event or incorrect treatment decisions that can result in patient injury, including medical intervention to prevent permanent injury or impairment. Additionally, due to the unknown storage and handling and storage conditions these sensors may have undergone during the distribution, these sensors may present additional risk. Lot Code: UDI-DI: 00386270004109 Lot Numbers: 1725204004 and 1725069002 Distribution: Dexcom does not have distribution records for this product due to its unauthorized release. However Dexcom is able to identify sensor session starts for the affected lots through the Dexcom G7 CGM App user performance data. Based on App user performance data, Lot 1725204004 has sensor session starts in the United States. Based on App user performance data, Lot 1725069002 has sensor session starts in the United States and the countries of Australia, Austria, Canada, Czech Republic, Finland, France, Germany, Hong Kong, Israel, Italy, Japan, South Korea, Kuwait, Netherlands, Poland, Portugal, Qatar, Singapore, Spain, Switzerland, Turkey, United Arab Emirates,and the United Kingdom. Affected quantity: N/A
- Hazard
- Certain lots of Dexcom G7 sensors originally designated as scrap and intended for destruction were stolen during the destruction process. The sensors were later sold to pharmacies and distributors by Pharmsource, LLC. Lot 1725204004 were designated for destruction because the G7 sensors were not properly sterilized and have an increased risk of skin infection if used. Lot 1725069002 were designated for destruction due to an increased risk of having no sensor readings available or experiencing sensor inaccuracy that includes the potential for missed detection of a hyperglycemic event or incorrect treatment decisions that can result in patient injury, including medical intervention to prevent permanent injury or impairment. Additionally, due to the unknown storage and handling and storage conditions these sensors may have undergone during the distribution, these sensors may present additional risk.