Philips Ingenuity CT scanners recalled for possible rescan risk
Philips Medical Systems is recalling certain Ingenuity CT scanner models. The clinical CT scan may not automatically start after the Bolus Tracking scan and may need to be triggered manually. This may lead to extra Bolus Tracking scans, clinical images that are inadequate for diagnosis, and a possible need to rescan the patient, which would add radiation exposure.
From the FDA notice · Read the official notice ↗ · All Philips recalls
What to do
- Check which Ingenuity CT systems at your facility match the catalog numbers (FCO72800815, FCO72800816, 728321, 728323, 728324, 728325, 728326, 728327) or the UDI numbers in the notice.
- The notice doesn't say how to get a fix; contact Philips Medical Systems about the affected systems and any corrective steps.
Is yours affected?
Compare your label or packaging with the recalled product.
- Product
- Ingenuity CT CT5000 Ingenuity Plus CT5000 Ingenuity Pro CT5000 Ingenuity Premium Ingenuity CT Upgrades Ingenuity CT China Ingenuity Core 128 Elite Brazil
- Sold in
- Nationwide
- Made for
- All ages (under 100 years)
This recall covers 3 products — see all of them.
Details
- Units affected
- 182
- Illnesses reported
- None reported
- Manufactured in
- 222 Jacobs St, N/A, Cambridge, MA, United States
- Classification
- Class II
- Recall number
- Z-2652-2026
In the FDA's words
- Description
- Ingenuity CT CT5000 Ingenuity Plus CT5000 Ingenuity Pro CT5000 Ingenuity Premium Ingenuity CT Upgrades Ingenuity CT China Ingenuity Core 128 Elite Brazil The clinical CT scan may not automatically trigger after the Bolus Tracking scan and may need to be triggered manually. Additional Bolus Tracking shots may occur. This may result in the clinical acquisition being inadequate for diagnosis, causing a potential for patient rescan and receiving additional radiation. FCO72800815, FCO72800816, 728321 DI (01)00884838059498 728323 DI (01)00884838059504 728326 DI (01)00884838059511 728327 DI (01)00884838098701 728324 DI (01)00884838059863 728325 Distribution: US Nationwide Distribution including Puerto Rico Affected quantity: 182
- Hazard
- The clinical CT scan may not automatically trigger after the Bolus Tracking scan and may need to be triggered manually. Additional Bolus Tracking shots may occur. This may result in the clinical acquisition being inadequate for diagnosis, causing a potential for patient rescan and receiving additional radiation.
3 products in this recall
Philips CT scanners recalled for possible failed automatic scan triggering
Philips is recalling 3 CT scanner product lines: Brilliance iCT (including Brilliance CT 64 Channel and CT6000 upgrades), IQon Spectral CT (including its upgrades), and Ingenuity CT (including the CT5000, Ingenuity Plus, Pro, Premium, Core 128 Elite, and upgrade versions). The clinical CT scan may not automatically start after the Bolus Tracking scan and may need to be triggered manually, and extra Bolus Tracking scans may occur. This could make the clinical images inadequate for diagnosis, possibly requiring a patient rescan and additional radiation exposure.
The agency lists each product separately. Check every one you might have.
- Brilliance iCT Brilliance iCT Upgrades Brilliance iCT SP Brilliance CT 64 Channel CT6000 Brilliance 64 UpgradesZ-2650-2026 · 128
- IQon Spectral CT IQon Spectral CT UpgradesZ-2651-2026 · 42
- Ingenuity CT CT5000 Ingenuity Plus CT5000 Ingenuity Pro CT5000 Ingenuity Premium Ingenuity CT Upgrades Ingenuity CT China Ingenuity Core 128 Elite Brazil (this page)Z-2652-2026 · 182