Moderate riskFDA DeviceClass II

Elekta brachytherapy applicator recalled for tip detachment risk

Elekta, Inc. is recalling 11 lots of its Advanced Gynecological Applicator, a brachytherapy device used for intracavitary cancer treatment of the cervix and endometrium. The recall is due to safety concerns that the tip of the intrauterine (IU) tube may detach and remain within a patient. The notice reports no injuries.

From the FDA notice · Read the official notice ↗ · All Elekta recalls

What to do

  1. Set aside affected lots and do not use them on patients.
  2. Check device lot codes against the 11 listed in the notice, starting with 08717213055001, 08717213055018 and 08717213055025.
  3. The notice does not specify a remedy; contact Elekta, Inc. for instructions on returning or replacing affected devices.

Is yours affected?

Compare your label or packaging with the recalled product.

Product
Brand Name: Brachytherapy Applicator Product Name: Advanced Gynecological Applicator Product Description: The intended use of the Advanced Gynecological Applicator is intracavitary brachytherapy for cancer treatment of the cervix and endometrium. Optional needles can be placed for interstitial brachytherapy.
Sold in
Nationwide
Made for
All ages (0 months and older)

This recall covers 3 products — see all of them.

Details

Units affected
586
Illnesses reported
None reported
Manufactured in
400 Perimeter Center Ter NE Ste 50, N/A, Atlanta, GA, United States
Classification
Class II
Recall number
Z-2711-2026

In the FDA's words

Description
Brand Name: Brachytherapy Applicator Product Name: Advanced Gynecological Applicator Product Description: The intended use of the Advanced Gynecological Applicator is intracavitary brachytherapy for cancer treatment of the cervix and endometrium. Optional needles can be placed for interstitial brachytherapy. Due to safety concerns regarding the detachment of the tip of the intrauterine tube (IU tube) that may remain within a patient. Lot Code: 08717213055001, 08717213055018, 08717213055025, 08717213055049, 08717213055056, 08717213055063, 08717213055087, 08717213055094, 08717213055100, 08717213055117, 08717213055131 Distribution: Worldwide - U.S. Nationwide distribution including in the states of AR, AZ, CA, CO, CT, DC, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MO, MS, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, SC, SD, TX, UT, VA, WI, and WV. The countries of Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bolivia, Botswana, Brazil, Burkina Faso, Cambodia, Canada, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, Georgia, Germany, Greece, Guatemala, Honduras, Hong Kong, Hungary, India, Indonesia, Iran (Islam Republic of), Iraq, Ireland, Italy, Japan, Kazakhstan, Kenya, Korea (Republic of), Kosovo, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Norway, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Romania, Russian Federation, Rwanda, Saudi Arabia, Singapore, Slovakia, South Africa, Spain, Swaziland, Sweden, Switzerland, Taiwan, Thailand, Togo, Turkey, Ukraine, United Arab Emirates, United Kingdom, Venezuela, Viet Nam, and Zambia. Affected quantity: 586
Hazard
Due to safety concerns regarding the detachment of the tip of the intrauterine tube (IU tube) that may remain within a patient.

3 products in this recall

Fletcher, Geneva and Brachytherapy applicators recalled for tip detachment

Elekta is recalling three intracavitary brachytherapy applicators used to treat cervical and endometrial cancer: the Fletcher CT/MR Shielded Applicator Set, the Brachytherapy Applicator (Advanced Gynecological Applicator), and the Geneva Brachytherapy Applicator. The intrauterine tube tip may detach and remain inside a patient. The notice reports no injuries.