Biomerieux VITEK 2 AST-N447 kits recalled for false resistance risk
Biomerieux is recalling one lot of VITEK 2 AST-N447 test kits, product reference 424620, because the kits may give false resistant results for carbapenem and/or cephalosporin antibiotics when pouch foil is fractured. The affected lot includes 45,900 cards distributed to Belgium and the Netherlands, and the product is not registered in the US (outside the US only). No injuries have been reported.
From the FDA notice · Read the official notice ↗ · All bioMérieux recalls
What to do
- Stop using kits from lot 0853293604 (UDI-DI 03573026634551) and set aside any remaining cards from that lot.
- Check whether any pouches show foil fractures, since the risk is linked to kits with fractured pouch foil.
- The notice does not specify a remedy; contact Biomerieux for instructions on the affected lot.
Is yours affected?
Compare your label or packaging with the recalled product.
- Product
- VITEK 2 AST-N447 TEST KIT Product reference 424620
- Sold in
- Nationwide
- Made for
- All ages (0 months and older)
Details
- Units affected
- 45,900 cards
- Illnesses reported
- None reported
- Manufactured in
- 595 Anglum Rd, N/A, Hazelwood, MO, United States
- Classification
- Class II
- Recall number
- Z-2715-2026
In the FDA's words
- Description
- VITEK 2 AST-N447 TEST KIT Product reference 424620 Potential for false resistant carbapenem and/or cephalosporin results when analyzing samples with kits from affected lot that have pouch foil fractures. Lot 0853293604 UDI-DI 03573026634551 Product is not registered in the US. OUS only. Distribution: International distribution to the countries of Belgium and the Netherlands. Affected quantity: 45,900 cards
- Hazard
- Potential for false resistant carbapenem and/or cephalosporin results when analyzing samples with kits from affected lot that have pouch foil fractures.