Think Surgical TMINI arrays recalled for possible misalignment
The knob on some TMINI tracking arrays may loosen, which could cause the array to become misaligned. The arrays are used with the TMINI Miniature Robotic System to guide knee replacement surgery. No injuries have been reported.
From the FDA notice · Read the official notice ↗
What to do
- Set aside affected tracking arrays and do not use them in knee replacement procedures.
- Check for REF 109003 (femur, UDI-DI M9461090030) and REF 109004 (tibia, UDI-DI M9461090040) arrays.
- Contact Think Surgical, Inc. for next steps, because the notice does not specify a remedy.
Is yours affected?
Compare your label or packaging with the recalled product.
- Product
- TMINI Femur Tracking Array REF 109003 UDI-DI code: M9461090030 TMINI Tibia Tracking Array REF 109004 UDI-DI code: M9461090040 TMINI Miniature Robotic System is indicated as a stereotaxic instrumentation system for total knee replacement (TKA) surgery. It is to assist the surgeon by providing software-defined spatial boundaries for orientation and reference information to identifiable anatomical structures for the accurate placement of knee implant components
- Sold in
- Nationwide
- Made for
- Adults (0 months and older)
Details
- Units affected
- 360 devices
- Illnesses reported
- None reported
- Manufactured in
- 47201 Lakeview Blvd, N/A, Fremont, CA, United States
- Classification
- Class II
- Recall number
- Z-2716-2026
In the FDA's words
- Description
- TMINI Femur Tracking Array REF 109003 UDI-DI code: M9461090030 TMINI Tibia Tracking Array REF 109004 UDI-DI code: M9461090040 TMINI Miniature Robotic System is indicated as a stereotaxic instrumentation system for total knee replacement (TKA) surgery. It is to assist the surgeon by providing software-defined spatial boundaries for orientation and reference information to identifiable anatomical structures for the accurate placement of knee implant components Due to manufacturing/design issue where the knob on the tracking array may loosen causing potential misalignment of the array. Model Number: 109003 UDI-DI code: M9461090030 Model Number: 109004 UDI-DI code: M9461090040 Distribution: U.S. Nationwide distribution in the states of AZ, CA, DE, FL, IA, KS, LA, MI, MN, MO, MS, NC, NE, NJ, NV, NY, OH, OR, PA, TX, UT, VA, and WA. Affected quantity: 360 devices
- Hazard
- Due to manufacturing/design issue where the knob on the tracking array may loosen causing potential misalignment of the array.