Medical Action angiography kits recalled for possible particulate
Medical Action Industries is recalling Pack Angiography NCMC 3/Cs BANNER kits (REF NCAG62I) because they contain Medline Namic Angiographic Star Off Handle Manifolds, which are the subject of another recall. The manifolds may have particulate in the fluid path, which could enter blood circulation and become lodged in blood vessels. The FDA has classified this as Class I, and 657 units were distributed in Arizona and Colorado.
From the FDA notice · Read the official notice ↗ · All Medical Action Industries recalls
What to do
- Set aside kits with REF NCAG62I and lot numbers 342513, 336758 or 338977 (UDI-DI 20809160490416) and do not use them.
- Contact Medical Action Industries for instructions; the notice does not specify a remedy.
Is yours affected?
Compare your label or packaging with the recalled product.
- Product
- Pack Angiography NCMC 3/Cs BANNER, REF: NCAG62I, containing Medline Namic Angiographic Star Off Handle Manifolds
- Sold in
- US: AZ, CO
- Made for
- All ages (under 100 years)
Details
- Units affected
- 657
- Illnesses reported
- None reported
- Manufactured in
- 25 Heywood Rd, N/A, Arden, NC, United States
- Classification
- Class I
- Recall number
- Z-2792-2026
In the FDA's words
- Description
- Pack Angiography NCMC 3/Cs BANNER, REF: NCAG62I, containing Medline Namic Angiographic Star Off Handle Manifolds Cardiovascular procedure kits contain manifolds, the subject of another recall, that may have particulate within the fluid path, which has the potential to be introduced into blood circulation and become lodged within blood vessels. UDI-DI: 20809160490416. Lot/Expiration: 342513(1/16/2028), 336758(9/20/2027), 338977(1/16/2028) Distribution: US: AZ, CO Affected quantity: 657
- Hazard
- Cardiovascular procedure kits contain manifolds, the subject of another recall, that may have particulate within the fluid path, which has the potential to be introduced into blood circulation and become lodged within blood vessels.