Halyard Cardiac Cath Pack kits recalled for particulate risk
Halyard Cardiac Cath Pack kits (EAMC1000-05) contain Medline Namic Angiographic Star Off Handle Manifolds that were recalled by the manufacturer. The manifolds may have particulates in their fluid path, which could enter the bloodstream and lodge in blood vessels. The recall covers 5,168 kits distributed nationwide, with the notice listing distribution in NC and TX.
From the FDA notice · Read the official notice ↗ · All AVID Medical recalls
What to do
- Set aside any Halyard Cardiac Cath Pack kits (EAMC1000-05) with the affected lot numbers listed in the notice.
- Check the UDI-DI 10809160377314 and lot numbers 1573593, 1583827, 1609541, 1612134, 1612516, 1619132, 1623384, 1627013, 1635459, 1672306, or 1675159.
- Contact AVID Medical, Inc. for next steps, since the notice does not specify a remedy.
Is yours affected?
Compare your label or packaging with the recalled product.
- Product
- Halyard convenience kits containing recalled Medline Namic Angiographic Star Off Handle Manifolds. CARDIAC CATH PACK EAMC1000-05
- Sold in
- Nationwide
- Made for
- All ages (under 120 years)
This recall covers 3 products — see all of them.
Details
- Units affected
- 5168 kits
- Illnesses reported
- None reported
- Manufactured in
- 9000 Westmont Dr, Toano, VA, United States
- Classification
- Class I
- Recall number
- Z-2796-2026
In the FDA's words
- Description
- Halyard convenience kits containing recalled Medline Namic Angiographic Star Off Handle Manifolds. CARDIAC CATH PACK EAMC1000-05 Affected kits contain Medline Namic Angiographic Star Off Handle Manifolds, which were recalled by the manufacturer due to presence of particulates within fluid path of the Manifolds. Particulate within the fluid path has the potential to be introduced into blood circulation and become lodged within blood vessels. UDI-DI 10809160377314 Lots 1573593 1583827 1609541 1612134 1612516 1619132 1623384 1627013 1635459 1672306 1675159 Distribution: US Nationwide distribution in the states of NC and TX. Affected quantity: 5168 kits
- Hazard
- Affected kits contain Medline Namic Angiographic Star Off Handle Manifolds, which were recalled by the manufacturer due to presence of particulates within fluid path of the Manifolds. Particulate within the fluid path has the potential to be introduced into blood circulation and become lodged within blood vessels.
3 products in this recall
Halyard cath kits with Medline manifolds recalled for particulates
AVID Medical is recalling three Halyard convenience kits that contain Medline Namic Angiographic Star Off Handle Manifolds, which the manufacturer recalled because of particulates in the fluid path. The particulates could enter the bloodstream and become lodged in blood vessels. The notice lists about 5,168 affected kits for each of the three products, with distribution in the US, including NC and TX.
The agency lists each product separately. Check every one you might have.
- Halyard convenience kits containing recalled Medline Namic Angiographic Star Off Handle Manifolds. CARDIAC CATH PACK EAMC1000-05 (this page)Z-2796-2026 · 5168 kits
- Halyard convenience kits containing recalled Medline Namic Angiographic Star Off Handle Manifolds. SAMMC ANGIOGRAPHY SAMM066-15Z-2797-2026 · 5168 kits
- Halyard convenience kits containing recalled Medline Namic Angiographic Star Off Handle Manifolds. HEART CATH, SELF REGIONAL SELF131-05Z-2798-2026 · 5168 kits
More from AVID Medical
See all- Halyard Arthroscopy Pack recalled for mislabeled shelf label
ModerateFDANov 5, 2026
- Halyard MINOR PACK kits recalled for potential open seals
ModerateFDANov 3, 2026
- Halyard ANTERIOR HIP PACK (SOAH31) recalled for non-sterile component
ModerateFDAOct 22, 2026
- AVID Medical DAVINCI TRUCUSTOM BUNDLE recalled for manufacturing lubricant contamination
ModerateFDAOct 14, 2026
- Halyard IV START KIT NON-STERILE recalled for open or incomplete applicator seal
ModerateFDASep 30, 2026