Permobil M5 Corpus power wheelchairs recalled for possible instability
Permobil is recalling M5 Corpus power wheelchairs because the Power Platform firmware may be missing drive inhibit logic. This may allow the chair to drive under restricted conditions and may create an unstable condition in certain circumstances. The recall is being reported retroactively, and no injuries have been reported.
From the FDA notice · Read the official notice ↗
What to do
- Check whether your chair's Power Platform firmware is version 1.27 through 1.57 (UDI-DI 17330818334562).
- Contact Permobil and your care provider before continuing to use an affected chair; the notice doesn't say how to get a fix or replacement.
Is yours affected?
Compare your label or packaging with the recalled product.
- Product
- M5 Corpus Power Wheelchair REF,120032
- Sold in
- Nationwide
- Made for
- All ages (18 years to 100 years)
Details
- Units affected
- 715 wheelchairs
- Illnesses reported
- None reported
- Manufactured in
- 300 Duke Dr, N/A, Lebanon, TN, United States
- Classification
- Class II
- Recall number
- Z-2805-2026
In the FDA's words
- Description
- M5 Corpus Power Wheelchair REF,120032 This recall is being reported retroactively. Power Platform firmware is missing specific drive inhibit logic. This issue may allow the chair to drive under restricted conditions, and may create an unstable condition in certain circumstances. Power Platform firmware versions 1.27 through 1.57 UDI-DI 17330818334562 Distribution: Worldwide - US: Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KY, ME, MI, MN, MO, MS, MT, NC, ND, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI, WV, WY and the country of CANADA. Affected quantity: 715 wheelchairs
- Hazard
- This recall is being reported retroactively. Power Platform firmware is missing specific drive inhibit logic. This issue may allow the chair to drive under restricted conditions, and may create an unstable condition in certain circumstances.