Low riskFDA DeviceClass III

Syva Emit Disopyramide calibrator recalled for azide concentration

Lot U1 of the Syva Emit Disopyramide Enzyme Immunoassay calibrator has an out-of-specification sodium azide concentration. The notice says assay results were not affected, and the health risk to lab technicians is low, with the worst outcome being mild skin or eye irritation. No injuries have been reported.

From the FDA notice · Read the official notice ↗ · All Siemens recalls

What to do

  1. Set aside Syva Emit Disopyramide Calibrator Lot U1 and don't use it.
  2. Check for UDI-DI 00842768001406 with Lot U1 on your calibrator kits.
  3. The notice doesn't say how to get a replacement; contact Siemens Healthcare Diagnostics.

Is yours affected?

Compare your label or packaging with the recalled product.

Product
Syva Emit Disopyramide Assay Enzyme Immunoassay, Lidocaine SMN 10445370 Catalog No 6N319UL
Sold in
Nationwide
Made for
All ages (0 months and older)

Details

Units affected
8 US; 44 OUS
Illnesses reported
None reported
Manufactured in
500 Gbc Dr, Newark, DE, United States
Classification
Class III
Recall number
Z-2880-2026

In the FDA's words

Description
Syva Emit Disopyramide Assay Enzyme Immunoassay, Lidocaine SMN 10445370 Catalog No 6N319UL Out-of-specification sodium azide concentration in Syva Emit Disopyramide Enzyme Immunoassay Calibrator Lot U1. While the assay results were not impacted, the overall health risk to lab technicians due to this issue is low, as exposure is limited by small volumes and a 0.2% post-reconstitution sodium azide concentration, with the worst outcome being mild skin or eye irritation and no serious harm anticipated. UDI-DI 00842768001406 Lot U1 Distribution: Worldwide - US Nationwide distribution in the states of CA and VA. The country of Japan. Affected quantity: 8 US; 44 OUS
Hazard
Out-of-specification sodium azide concentration in Syva Emit Disopyramide Enzyme Immunoassay Calibrator Lot U1. While the assay results were not impacted, the overall health risk to lab technicians due to this issue is low, as exposure is limited by small volumes and a 0.2% post-reconstitution sodium azide concentration, with the worst outcome being mild skin or eye irritation and no serious harm anticipated.

More from Siemens

See all

Related topics

Industrial chemicalsOtherMedical devicesSold nationwideMore from FDA Device