Moderate riskFDA DeviceClass II

United Orthopedic U2 Total Knee adaptor recalled over screw defect

United Orthopedic Corp. is recalling two U2 Total Knee System femoral valgus adaptors (left and right). A subcontractor did not fully complete the process on a component screw, which can prevent the stem trail from being assembled properly. The recall covers model 9323-5333 and 9323-5334 and was distributed in Texas and Michigan. No injuries have been reported.

From the FDA notice · Read the official notice ↗

What to do

  1. Check whether your facility has model 9323-5333 (lot 25FB50) or 9323-5334 (lot 25FB51) femoral valgus adaptors.
  2. Do not use the affected adaptors until you have confirmed the next steps with the manufacturer.
  3. Contact United Orthopedic Corp. for instructions on returning or replacing affected units.

Is yours affected?

Compare your label or packaging with the recalled product.

Product
Brand Name: U2 Total Knee System Product Name: Femoral valgus adaptor, left and right Model/Catalog Number: 9323-5333 and 9323-5334 Software Version: N/A Product Description: Manual Orthopedic Surgical Instrument Component: No
Sold in
Nationwide
Made for
All ages (0 months and older)

This recall covers 2 products — see all of them.

Details

Units affected
2
Illnesses reported
None reported
Manufactured in
No. 16 Luke 1st Rd., Luzhu Dist., Kaohsiung, Taiwan
Classification
Class II
Recall number
Z-2885-2026

In the FDA's words

Description
Brand Name: U2 Total Knee System Product Name: Femoral valgus adaptor, left and right Model/Catalog Number: 9323-5333 and 9323-5334 Software Version: N/A Product Description: Manual Orthopedic Surgical Instrument Component: No Due to an incomplete process of the component screw by a subcontractor which prevents the properly assembling of stem trail. Product Number; UDI-DI Code; Lot Code: Product number 9323-5333; UDI-DI 04719886908084; Lot number 25FB50 Product number 9323-5334; UDI-DI 04719886908091; Lot number 25FB51 Distribution: US Nationwide distribution in the states of Texas and Michigan. Affected quantity: 2
Hazard
Due to an incomplete process of the component screw by a subcontractor which prevents the properly assembling of stem trail.

2 products in this recall

U2 Total Knee System trial parts recalled for assembly defect

United Orthopedic is recalling two U2 Total Knee System surgical trial parts: the PSA Box Trial (left and right, sizes 1-6) and the Femoral valgus adaptor (left and right). A subcontractor did not fully finish a screw, so the trial stem cannot be assembled properly. The products were distributed nationwide, with confirmed distribution in Texas and Michigan. No injuries have been reported.

The agency lists each product separately. Check every one you might have.

More from United Orthopedic

Related topics