Beckman Coulter MicroScan software recalled for possible false results
Beckman Coulter is recalling 158 MicroScan instruments with LabPro V5.03 Panel Update 11 installed. The analyzers contain incorrect algorithms that may affect antimicrobial susceptibility test results, leading to false susceptibility classifications. No injuries have been reported.
From the FDA notice · Read the official notice ↗ · All Beckman Coulter recalls
What to do
- Check whether your instrument has LabPro V5.03 Panel Update 11 installed (UDI 15099590799243 / V5.03 Panel Update 11) on model B1018-280, B1018-283, B1018-284, B1018-440, or B1018-496.
- Be aware that antimicrobial susceptibility results from these instruments may be affected.
- The notice doesn't specify a remedy; contact Beckman Coulter for instructions.
Is yours affected?
Compare your label or packaging with the recalled product.
- Product
- REF # D23352, MicroScan LabPro V5.03 Panel Update 11 (US), Electronic installed on the following instruments: - B1018-280 MicroScan autoSCAN-4 Instrument - B1018-283 MicroScan WalkAway-40 plus Instrument - B1018-284 MicroScan WalkAway-96 plus Instrument - B1018-440 DxM-1040 MicroScan WalkAway Instrument - B1018-496 DxM-1096 MicroScan WalkAway Instrument
- Sold in
- Nationwide
- Made for
- All ages (under 100 years)
This recall covers 2 products — see all of them.
Details
- Units affected
- 158 instruments
- Illnesses reported
- None reported
- Manufactured in
- 250 S Kraemer Blvd, Brea, CA, United States
- Classification
- Class II
- Recall number
- Z-2888-2026
In the FDA's words
- Description
- REF # D23352, MicroScan LabPro V5.03 Panel Update 11 (US), Electronic installed on the following instruments: - B1018-280 MicroScan autoSCAN-4 Instrument - B1018-283 MicroScan WalkAway-40 plus Instrument - B1018-284 MicroScan WalkAway-96 plus Instrument - B1018-440 DxM-1040 MicroScan WalkAway Instrument - B1018-496 DxM-1096 MicroScan WalkAway Instrument Analyzers contain incorrect algorithms which may affect antimicrobial susceptibility test results, leading to false susceptibility classifications." UDI: 15099590799243/ V5.03 Panel Update 11 Distribution: US Nationwide distribution including in the states of AK, AL, AR, CA, CO, DE, FL, GA, IA, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, VA, WI, WV, WY. Affected quantity: 158 instruments
- Hazard
- Analyzers contain incorrect algorithms which may affect antimicrobial susceptibility test results, leading to false susceptibility classifications."
2 products in this recall
MicroScan LabPro software recalled for possible incorrect results
Beckman Coulter is recalling two MicroScan LabPro V5.03 Panel Update 11 products, an electronic update (REF D23352) and a DVD kit (REF D23354), for installation on MicroScan instruments such as the autoSCAN-4 and WalkAway-40 plus. The analyzers contain incorrect algorithms that may affect antimicrobial susceptibility test results and lead to false susceptibility classifications. The recall covers products distributed nationwide, including the listed states.
The agency lists each product separately. Check every one you might have.
- REF # D23352, MicroScan LabPro V5.03 Panel Update 11 (US), Electronic installed on the following instruments: - B1018-280 MicroScan autoSCAN-4 Instrument… (this page)Z-2888-2026 · 158 instruments
- REF # D23354, MicroScan LabPro V5.03 Panel Update 11 DVD Kit (US) on the following instruments: - B1018-280 MicroScan autoSCAN-4 Instrument - B1018-283 Mi…Z-2889-2026