Moderate riskFDA DeviceClass II

Remel Neisseria meningitidis Gp C test recalled for poor performance

Certain lots of the Remel (Thermo Fisher Diagnostics) Neisseria meningitidis Group C 2 mL test, product R30166901, did not perform as its instructions require. The test did not return a positive result with a reference control strain within 60 seconds. About 26 units were distributed nationwide, and no injuries have been reported.

From the FDA notice · Read the official notice ↗ · All Remel recalls

What to do

  1. Check whether your lab has product R30166901 with lot 6214791, 6175327, 6148809, or 6002010.
  2. Stop using any affected lots and set them aside until you receive guidance from the manufacturer.
  3. Contact Remel or Thermo Fisher Diagnostics for next steps, since the notice does not list a specific remedy.

Is yours affected?

Compare your label or packaging with the recalled product.

Product
Thermo Fisher Diagnostics Neisseria meningitidis Gp C 2ml/VL, Product Number R30166901; invitro diagnostics
Sold in
Nationwide
Made for
All ages (under 100 years)

Details

Units affected
26 units
Illnesses reported
None reported
Manufactured in
Clipper Boulevard West, Ken, Dartford, United Kingdom
Classification
Class II
Recall number
Z-2927-2026

In the FDA's words

Description
Thermo Fisher Diagnostics Neisseria meningitidis Gp C 2ml/VL, Product Number R30166901; invitro diagnostics Certain lots of Neisseria meningitidis Gp C 2ml/VL, R30166901, are not performing to IFU criteria not positive to ATCC13102 within 60 Seconds. UDI/DI 05056080501161, Lot Numbers: 6214791, exp. 18/03/2025; 6175327, exp. 04/12/2027; 6148809, exp. 30/09/2027; 6002010, exp. 31/07/2027. Distribution: US Nationwide distribution. Affected quantity: 26 units
Hazard
Certain lots of Neisseria meningitidis Gp C 2ml/VL, R30166901, are not performing to IFU criteria not positive to ATCC13102 within 60 Seconds.

Related topics