Moderate riskFDA DeviceClass II

Bioporto ProNephro AKI kit recalled for unvalidated reagent rework

Bioporto Diagnostics recalled certain lots of the ProNephro AKI (NGAL) Reagent Kit, a test used to help detect acute kidney injury. The reagents were reworked using a process that had not been validated, which could lead to missed kidney disease, incorrect treatment decisions, or inappropriate monitoring. No injuries have been reported.

From the FDA notice · Read the official notice ↗

What to do

  1. Check whether your kit's model number ST101UB and lot number 04909125 or 04912125 match the recall.
  2. Stop using kits from the affected lots.
  3. Contact Bioporto Diagnostics or your supplier for return and remedy instructions.
  4. Talk to your healthcare provider if results from affected lots were used in patient care decisions.

Is yours affected?

Compare your label or packaging with the recalled product.

Product
ProNephro AKI (NGAL) Reagent Kit (100 tests). IVD Acute Kidney Injury Test System.
Sold in
Nationwide
Made for
All ages (0 months and older)

Details

Units affected
400 units
Illnesses reported
None reported
Manufactured in
Tuborg Havnevej 15, St, Hellerup, Denmark
Classification
Class II
Recall number
Z-2937-2026

In the FDA's words

Description
ProNephro AKI (NGAL) Reagent Kit (100 tests). IVD Acute Kidney Injury Test System. The use of a non-validated rework process for assay reagents which could lead to missed diagnosis of renal disease, inappropriate treatment decisions, or inappropriate monitoring of renal disease in intended users. Model number: ST101UB / 08787174 190; GTIN code: 5711435000156; Lot-number: 04909125 and 04912125. Distribution: Worldwide - The affected product distribution included the United States and the countries of European Union, Chile, and Hong Kong. No affected product was distributed to Canada; however, Health Canada was notified because the product is registered in Canada. Affected quantity: 400 units
Hazard
The use of a non-validated rework process for assay reagents which could lead to missed diagnosis of renal disease, inappropriate treatment decisions, or inappropriate monitoring of renal disease in intended users.