FDA-Device2026-08-26Class IIPROCESSING DEFECT

Spectral Instruments Ami HTX X-ray system recalled for missing on-screen safety indicator

OTHERNationwide distributionMedical devices

What this means

The Ami HTX X-ray system may briefly fail to display the "X-Ray On" banner during the run readiness check. This missing visual indicator could create confusion about whether the X-ray is active. No injuries have been reported.

  • If you operate an Ami HTX system, contact Spectral Instruments Inc for instructions or a firmware update
  • Verify the X-ray status through alternative confirmation methods until the issue is resolved
Get recalls like this in a free plain-English brief every Sunday.Get alerts ↓
The agency's original wording
Hazard

Momentary absence of On-Screen "X-Ray On" banner during run readiness check.

Intended for
All ages
Illnesses reported
None reported
Sold states
United States.
Affected count
35
Manufactured in
420 N Bonita Ave, Tucson, AZ, United States
Products
Ami HTX

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2940-2026